Enhancing blood culture volume with ultrathin-wall cannula devices
Authors:
Open Access
- 8 October 2025
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 63 (10) , e0020825
Abstract
Suboptimal blood volume collection negatively affects pathogen recovery and time to pathogen detection in clinical blood culture samples. This study evaluated the performance of the ultrathin-wall cannula device of the BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch; BD Life Sciences, Franklin Lakes, NJ) in increasing the blood volume collected compared to regular cannulas. In this retrospective study, the volume of blood collected in negative BD BACTEC Plus Aerobic/F Culture Vials prior to and following blind conversion to UltraTouch was measured from blood background metabolic activity signal acquired during the initial incubation period in the BD BACTEC FX instrument using a custom software application (patent pending). Differences in average fill volume between pre- and post-conversion were evaluated overall and by month. A separate logistic regression was also performed to model the relationship between monthly average blood volume increase (mL) and sample positivity rate. Prevalence rates for identified organisms were also assessed. The blood fill volume of 13,356 pre-conversion and 119,971 post-conversion bottles was compared. The average post-conversion fill volume was significantly higher than that of pre-conversion (7.56 mL [95% CI, 7.54–7.59] vs. 5.68 mL [95% CI, 5.62–5.74], respectively). This 1.88 mL increase in collected blood volume increased sample positivity from 7.0% to 7.7% ( P = 0.002; odds ratio [OR], 1.11; 95% CI, 1.04–1.19). This study demonstrated that using an ultrathin-wall cannula blood collection device can significantly increase blood culture volume and sample positivity, supporting improved identification of patients suffering from bloodstream infections. IMPORTANCE: Substandard blood fill volume practices in clinical settings negatively affect pathogen recovery which, in turn, may impact patients’ care. Whereas professional education regarding the importance of adhering to clinical guidelines has been helpful in partly remediating the issue, the role of blood collection set technologies in optimizing blood culture volume has not been studied as thoroughly. Our study assessed the performance of the BD Vacutainer UltraTouch Push Button Blood Collection Set, which is designed with an ultrathin-wall cannula, to improve blood culture volume.Keywords
This publication has 18 references indexed in Scilit:
- Blood culture quality assurance: findings from a RCPAQAP Key Incident Monitoring and Management Systems (KIMMS) audit of blood culture performancePathology, 2023
- Blood Culture Utilization in the Hospital Setting: a Call for Diagnostic StewardshipJournal of Clinical Microbiology, 2022
- Seasonal variations in blood culture numbers and time to positivity and potential impact of reducing incubation periodsEuropean Journal of Clinical Microbiology & Infectious Diseases, 2021
- Detailed Analysis of the Characteristics of Sample Volume in Blood Culture BottlesJournal of Clinical Microbiology, 2019
- Active Monitoring and Feedback to Improve Blood Culture Fill Volumes and Positivity Across a Large Integrated Health SystemClinical Infectious Diseases, 2019
- Establishing a Long-Term Model for Analysis and Improvement of Underfilled Blood Culture VolumesAmerican Journal of Clinical Pathology, 2018
- How to achieve accurate blood culture volumes: the BD BACTEC FX blood volume monitoring system as a measuring instrument and educational toolEuropean Journal of Clinical Microbiology & Infectious Diseases, 2018
- ggplot22016
- Effects of Volume and Site of Blood Draw on Blood Culture ResultsJournal of Clinical Microbiology, 2009
- Is the Volume of Blood Cultured Still a Significant Factor in the Diagnosis of Bloodstream Infections?Journal of Clinical Microbiology, 2007