Abstract
Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is widely used for therapeutic drug monitoring of immunosuppressants given to transplant recipients. This study evaluated the performance of the newly introduced MassTrak Immunosuppressant XE Kit (Waters Corporation; “the Kit”) in the determination of everolimus and cyclosporin A (CsA) using LC-MS/MS.The linearity, precision, detection limit, carryover and matrix effect of the Kit and comparison of the in-house method and Kit procedure were evaluated according to Clinical and Laboratory Standards Institute guidelines.The Kit afforded good linearity in the measurement of everolimus from 2 to 26 ng/mL (RThe Kit employing isotopically labeled internal standards provides reliable measurements of immunosuppressant levels over a broad range of concentrations.