Risk factors for femoral stem fracture following total hip arthroplasty: a systematic review and meta analysis
Open Access
- 12 April 2024
- journal article
- review article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 144 (5) , 2421-2428
Abstract
Background: Femoral stem fracture following total hip arthroplasty (THA) is an infrequent but nevertheless devastating complication, with an increasing worldwide prevalence as demand for primary THA continues to increase. The aim of this study was to perform a systematic review and meta-analysis of risk factors for femoral stem fracture to help identify at risk patients. Methods: A systematic search was conducted on EMBASE, MEDLINE and AMED to identify relevant studies. Data regarding study design, source, population, intervention, and outcomes was collated. Data extraction was performed on a custom form generated using Cochrane recommended methodology and analysis of risk factors performed including odds ratios (ORs) with 95% confidence intervals (CIs). Results: A total of 15 studies reporting a total of 402 stem fractures in 49 723 THAs were identified. The median time from index procedure to stem fracture was 68 months (IQR 42.5–118) whilst mean age at index surgery was 61.8 years (SD 6.9). Male gender (OR = 3.27, 95% CI = 2.59–4.13, p < 0.001), patient weight above 80 kg (OR = 3.55, 95% CI = 2.88–4.37, p < 0.001), age under 63 years (OR = 1.22, 95% CI = 1.01–1.49, p < 0.001), varus stem alignment (OR = 5.77, 95% CI = 3.83–8.7, p < 0.001), use of modular implants (OR = 1.95, 95% CI = 1.56–2.44, p < 0.01) and undergoing revision arthroplasty (OR = 3.33, 95% CI = 2.70–4.1, p < 0.001) were significant risk factors for prosthetic stem fracture. A risk window of 15 years post-surgery was identified. Conclusions: This review concludes that patient weight, younger age, male sex, varus stem alignment, revision arthroplasty and use of modular stems are significant risk factors for femoral stem fracture. Modifying these risk factors where possible may help reduce incidence of femoral stem fracture in at risk patients.Keywords
This publication has 29 references indexed in Scilit:
- Return to activity following revision total hip arthroplastyArchives of orthopaedic and trauma surgery, 2018
- What can the surgeon do to reduce the risk of junction breakage in modular revision stems?Arthroplasty Today, 2018
- Alarmingly High Rate of Implant Fractures in One Modular Femoral Stem Design: A Comparison of Two ImplantsThe Journal of Arthroplasty, 2017
- International survey of primary and revision total knee replacementInternational Orthopaedics, 2011
- Revision Total Hip Arthroplasty with a Porous-coated Modular Stem: 5 to 10 Years FollowupClinical Orthopaedics and Related Research, 2010
- Fractures of Modern High Nitrogen Stainless Steel Cemented StemsThe Journal of Arthroplasty, 2008
- Failure of the stem in total hip replacementArchives of orthopaedic and trauma surgery, 1988
- The mechanism of fracture of the femoral prosthesis in total hip replacementInternational Orthopaedics, 1979
- A comparative experimental study of stresses in femoral total hip replacement components: The effects of prostheses orientation and acrylic fixationJournal of Biomechanics, 1976
- Fatigue behavior of type 316 stainless steel under simulated body conditionsJournal of Biomedical Materials Research, 1971