Reflectance Thermometry for Microscale Metal Thin Films
- 1 May 2007
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 46 (5R) , 3117
Abstract
A noncontact temperature measurement system applicable to a sub-micrometer scale area based on a thermoreflectance technique has been developed. A high spatial resolution of 0.7 µm is achieved using a 785 nm near-infrared laser and a confocal microscope. A relative change in the reflectivity of the order of 10-6 is detected without any data accumulation. A temperature resolution of 0.2 °C at around 100 °C is achieved for a molybdenum thin strip sample of 70 nm thickness and 100 µm width. The simultaneous measurements of thermal variations in the relative intensity of the reflected light and in electrical resistance enable us to calibrate the temperature scale of thermoreflectance thermometry.Keywords
This publication has 4 references indexed in Scilit:
- High-resolution thermoreflectance microscopyMRS Proceedings, 2002
- Development of a thermal diffusivity measurement system for metal thin films using a picosecond thermoreflectance techniqueMeasurement Science and Technology, 2001
- Thermoreflectance measurements of transient temperature upon integrated circuits: application to thermal conductivity identificationMicroelectronics Journal, 1998
- Reflectivity of Metals at High TemperaturesJournal of Applied Physics, 1972