Performance evaluation of the Sysmex CN-6000-automated coagulation analyzers for routine coagulation tests
Authors:
- 22 December 2022
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Coagulation & Fibrinolysis
- Vol. 34 (2) , 93-98
Abstract
The Sysmex CN-6000 is a novel automated multiparameter coagulometer that performs clotting, chromogenic and immunological assays, and platelet aggregation tests in a single system. Here we evaluated its performance of routine coagulation assays. The precision, linearity, carryover and establishment of reference ranges of the CN-6000, as well as correlations between it and the currently used Diagnostica Stago STA-R Max were determined according to Clinical and Laboratory Standards Institute guidelines. The evaluated parameters included prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen (FBG), antithrombin (AT), d-dimers (DDi), and fibrin and FBG degradation products (FDP). The intra-run and inter-run precisions of the six tests were determined using normal and pathological control materials; all coefficients of variation were acceptable and within the allowable ranges. The CN-6000 showed excellent linearity for FBG, AT, DDi, and FDP (R = 0.999–1.00). Passing-Bablok regression (R 2 > 0.95) demonstrated good agreement between the analyzers. In the carryover study, APTT, PT, FBG, AT, DDi, and FDP values were all acceptable. The establishing reference intervals revealed that each manufacturer's range was acceptable. Significant differences were observed in the APTT reference range because of using different detection systems and reagents. The CN-6000 analyzer showed reliable performance and good correlation with the currently used STA-R Max automated hemostatic analyzer. As CN-6000 uses an optical clot-detection method, its reference ranges for PT and APTT are lower than those of the STA-R Max; thus, the difference should be considered before its use.Keywords
This publication has 11 references indexed in Scilit:
- International Council for Standardization in Haematology (ICSH) laboratory guidance for the verification of haemostasis analyser‐reagent test systems. Part 2: Specialist tests and calibrated assaysInternational Journal of Laboratory Hematology, 2021
- Analytical Performance and Comparability of Three New Coagulation AnalyzersClinical Laboratory, 2021
- International Council for Standardization in Haematology (ICSH) laboratory guidance for the evaluation of haemostasis analyser‐reagent test systems. Part 1: Instrument‐specific issues and commonly used coagulation screening testsInternational Journal of Laboratory Hematology, 2020
- A comparative evaluation of the CN‐6000 haemostasis analyser using coagulation, amidolytic, immuno‐turbidometric and light transmission aggregometry assaysInternational Journal of Laboratory Hematology, 2020
- Evaluation of a new thromboplastin reagent STA‐NeoPTimal on a STA R Max analyzer for the measurement of prothrombin time, international normalized ratio and extrinsic factor levelsInternational Journal of Laboratory Hematology, 2020
- Analytical Assessment of the New Roche Cobas t 711 Fully Automated Coagulation AnalyzerSeminars in Thrombosis and Hemostasis, 2019
- Validation of the Sysmex CS5100 coagulation analyzer and comparison to the Stago STA‐R analyzer for routine coagulation parametersInternational Journal of Laboratory Hematology, 2014
- Optical and Mechanical Clot Detection Methodologies: A Comparison Study for Routine Coagulation TestingJournal of Clinical Laboratory Analysis, 2012
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986