Validation of the professional device for blood pressure measurement Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization standard
Authors:
- Anastasios KolliasAnastasios Kollias
- Angeliki NtineriANAngeliki Ntineri
- Konstantinos G. KyriakoulisKKKonstantinos G. Kyriakoulis
- Emelina StambolliuESEmelina Stambolliu
- Styliani LagouSLStyliani Lagou
- Nadia BoubouchairopoulouNBNadia Boubouchairopoulou
- George S. StergiouGSGeorge S. Stergiou
- 1 April 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 23 (2) , 112-114
Abstract
To assess the accuracy of the professional oscillometric blood pressure (BP) monitor Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. Adults and children (aged 3–12 years) were recruited to fulfil the age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard using the same-arm sequential BP measurement method. Three cuffs of the test device were used for arm circumferences of 14–22, 22–32 and 32–42 cm. A total of 115 participants were recruited and 88 were included in the analysis (51 adults and 37 children). For criterion 1, the mean±SD of the differences between the test device and reference BP was −1.0±7.0/−4.7±5.4 mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference per participant was 5.15/4.77 mmHg (systolic/diastolic). The professional Microlife WatchBP Office BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard for adults and children and can be recommended for clinical use.Keywords
This publication has 6 references indexed in Scilit:
- Clinical Practice Guideline for Screening and Management of High Blood Pressure in Children and AdolescentsPediatrics, 2017
- Accuracy of Automated Blood Pressure Measurement in ChildrenHypertension, 2017
- Management of high blood pressure in children and adolescents: recommendations of the European Society of HypertensionJournal Of Hypertension, 2009
- Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocolBlood Pressure Monitoring, 2008
- Automated device that complies with current guidelines for office blood pressure measurement: design and pilot application study of the Microlife WatchBP Office deviceBlood Pressure Monitoring, 2008
- Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devicesJournal Of Hypertension, 1993