Stabilization of the Syndesmosis in the Maisonneuve Fracture—A Biomechanical Study Comparing 2-Hole Locking Plate and Quadricortical Screw Fixation
- 1 April 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 27 (4) , 212-216
Abstract
Objective: The aim of this study is to determine whether a 2-hole locking plate has biomechanical advantages over conventional screw stabilization of the syndesmosis in this injury pattern. Methods: Six pairs of fresh-frozen human cadaver lower legs were prepared to simulate an unstable Maisonneuve fracture. Each limb was compared with its pair; the syndesmosis in one being stabilized with two 4.5-mm quadricortical cortical screws, the other a 2-hole locking plate with 3.2-mm locking screws. The limbs were then mounted on a servohydraulic testing rig and axially loaded to a peak load of 800N for 12000 cycles. Fibula shortening and diastasis were measured. Each limb was then externally rotated until failure occurred. Failure was defined as fracture of bone or metalwork, syndesmotic widening, or axial migration >2 mm. Results: Both constructs effectively stabilized the syndesmosis during the cyclical loading within 0.1 mm of movement. However, the locking plate group demonstrated greater resistance to torque compared with quadricortical screw fixation (40.6 Nm vs. 21.2 Nm, respectively, P value < 0.03). Conclusion: A 2-hole locking plate (with 3.2-mm screws) provides significantly greater stability of the syndesmosis to torque when compared with 4.5-mm quadricortical fixation.Keywords
This publication has 11 references indexed in Scilit:
- Functional Outcomes After Syndesmotic Screw Fixation and RemovalJournal of Orthopaedic Trauma, 2010
- Injuries to the tibiofibular syndesmosisThe Journal of Bone and Joint Surgery. British volume, 2008
- Syndesmosis Fixation: Analysis of Shear Stress via Axial Load on 3.5-mm and 4.5-mm Quadricortical Syndesmotic ScrewsThe Journal of Foot and Ankle Surgery, 2006
- Distal tibiofibular syndesmosis fixation: A cadaveric, simulated fracture stabilization study comparing bioabsorbable and metallic single screw fixationThe Journal of Foot and Ankle Surgery, 2005
- Suture-Button Syndesmosis FixationClinical Orthopaedics and Related Research, 2005
- Screw fixation of the syndesmosis: a cadaver model comparing stainless steel and titanium screws and three and four cortical fixationInjury, 2005
- Biomechanics of Locked Plates and ScrewsJournal of Orthopaedic Trauma, 2004
- Biomechanical Comparison of Syndesmosis Fixation with 3.5- and 4.5-millimeter Stainless Steel ScrewsFoot & Ankle International, 2000
- The Effect of Fibular Malreduction on Contact Pressures in an Ankle Fracture Malunion Model*Journal of Bone and Joint Surgery, 1997
- Changes in tibiotalar area of contact caused by lateral talar shiftJournal of Bone and Joint Surgery, 1976