Pulse-contour derived cardiac output measurements in morbid obesity: influence of actual, ideal and adjusted bodyweight
Open Access
- 18 August 2017
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 32 (3) , 423-428
Abstract
The non-invasive Nexfin cardiac output (CO) monitor shows a low level of agreement with the gold standard thermodilution method in morbidly obese patients. Here we investigate whether this disagreement is related to excessive bodyweight, and can be improved when bodyweight derivatives are used instead. We performed offline analyses of cardiac output recordings of patient data previously used and partly published in an earlier study by our group. In 30 morbidly obese patients (BMI > 35 kg/m2) undergoing laparoscopic gastric bypass, cardiac output was simultaneously determined with PiCCO thermodilution and Nexfin pulse-contour method. We investigated if agreement of Nexfin-derived CO with thermodilution CO improved when ideal and adjusted—instead of actual- bodyweight were used as input to the Nexfin. Bodyweight correlated with the difference between Nexfin-derived and thermodilution-derived CO (r = −0.56; p = 0.001). Bland Altman analysis of agreement between Nexfin and thermodilution-derived CO revealed a bias of 0.4 ± 1.6 with limits of agreement (LOA) from −2.6 to 3.5 L min when actual bodyweight was used. Bias was −0.6 ± 1.4 and LOA ranged from −3.4 to 2.3 L min when ideal bodyweight was used. With adjusted bodyweight, bias improved to 0.04 ± 1.4 with LOA from −2.8 to 2.9 L min. Our study shows that agreement of the Nexfin-derived with invasive CO measurements in morbidly obese patients is influenced by body weight, suggesting that Nexfin CO measurements in patients with a BMI above 35 kg/m2 should be interpreted with caution. Using adjusted body weight in the Nexfin CO-trek algorithm reduced the bias.Keywords
This publication has 27 references indexed in Scilit:
- Health risk factors in the anesthesia populationJournal of Clinical Anesthesia, 2016
- Noninvasive Continuous Cardiac Output by the Nexfin Before and After Preload-Modifying ManeuversAnesthesia & Analgesia, 2013
- Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgeryBritish Journal of Anaesthesia, 2012
- Noninvasive continuous hemodynamic monitoringJournal of Clinical Monitoring and Computing, 2012
- Impact of non invasive and beat-to-beat arterial pressure monitoring on intraoperative hemodynamic managementJournal of Clinical Monitoring and Computing, 2012
- A comparison of the Nexfin ® and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgeryAnaesthesia, 2012
- Estimating Ideal Body Weight – A New FormulaObesity Surgery, 2005
- A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients †British Journal of Anaesthesia, 2001
- Measuring agreement in method comparison studiesStatistical Methods in Medical Research, 1999
- The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new modelJournal of Biomechanics, 1984