Torque loss under mechanical cycling of long‐span zirconia and titanium‐cemented and screw‐retained implant‐supported CAD/CAM frameworks
- 29 October 2013
- journal article
- research article
- Published by Wiley in Clinical Oral Implants Research
- Vol. 25 (12) , 1395-1402
Abstract
Objectives: This study evaluated the screw joint stability after cyclic loading of implant‐supported titanium and zirconia CAD/CAM frameworks for fixed dental prostheses (FDPs) with different retention methods.Material and methods: Twenty four one‐piece frameworks supported by six threaded implants placed in the maxilla were fabricated using a CAD/CAM technique (NeoShape). Dry‐pressed porcelain crowns were luted to the frameworks to standardize the specimens. The specimens were then divided into four groups (n = 6) according to framework material (titanium or zirconia) and retention method for the prosthesis (cement‐ or screw‐retained): G1, Ti‐cemented; G2, Ti‐screw‐retained; G3, Zr‐cemented; and G4, Zr‐screw‐retained. A digital torque ratchet was used to assess the initial preload removal torque. Torque was then reapplied and the specimens were submitted to a 200 N cyclic load, at a frequency of 2 Hz, underwater in controlled temperature of 37°, and for 1 × 106 cycles. An opposing lower dental arch was fabricated using bis‐acrylic resin to simulate occlusal contacts in centric. After cyclic loading, postload removal torque was measured. Preload and postload torque loss was expressed as a percentage of the initial load. Data were submitted to a linear mixed‐effects model for statistical significance (α = 0.05) to evaluate the effect of cyclic loading in the screw torque loss used with frameworks of different materials and retention methods.Results: Significant screw torque loss (%) was found for the tested groups (before/after cyclic loading, respectively): G1 (39.77/61.83), G2 (37.57/50.96), G3 (34.87/54.10), and G4 (47.56/73.50) (P < 0.05).Conclusions: The screw removal torque was significantly reduced for all groups in this study after cyclic loading the specimens. Screw‐retained zirconia specimens presented the highest torque loss before and after the cyclic loadings compared with the other specimens that were tested.Keywords
Funding Information
- FAPESP (2010/00124-3)
- FAPESP (2010/02218-5)
This publication has 33 references indexed in Scilit:
- Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crownsEuropean Journal of Oral Sciences, 2011
- Effects of Abutment Screw Coating on Implant PreloadJournal of Prosthodontics, 2010
- Retention of zirconia copings on zirconia implant abutments cemented with provisional luting agentsJournal of Oral Rehabilitation, 2009
- Edentulous Arch Treatment With a CAD/CAM Screw-Retained Framework and Cemented Crowns: A Clinical Case ReportJournal of Oral Implantology, 2009
- Y-TZP Ceramics: Key Concepts for Clinical ApplicationOperative Dentistry, 2009
- Fracture resistance of implant-supported screw- versus cement-retained porcelain fused to metal single crowns: SEM fractographic analysisDental Materials, 2007
- Abutment Rotational Displacement of External Hexagon Implant System Under Lateral Cyclic LoadingClinical Implant Dentistry and Related Research, 2006
- Inflammatory Infiltrate, Microvessel Density, Nitric Oxide Synthase Expression, Vascular Endothelial Growth Factor Expression, and Proliferative Activity in Peri‐Implant Soft Tissues Around Titanium and Zirconium Oxide Healing CapsJournal of Periodontology, 2006
- A systematic review of the survival and complication rates of fixed partial dentures (FPDs) after an observation period of at least 5 yearsClinical Oral Implants Research, 2004
- Fixture design and overload influence marginal bone loss and future success in the Brånemark® systemClinical Oral Implants Research, 1992