Acrylonitrile Butadiene Styrene (ABS) plastic‐based low cost tissue equivalent phantom for verification dosimetry in IMRT
Authors:
- Rajesh KumarRKRajesh Kumar
- V.S. ShaijuVSV.S. Shaiju
Open Access
- 17 December 2009
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 11 (1) , 24-32
Abstract
A novel IMRT phantom was designed and fabricated using Acrylonitrile Butadiene Styrene (ABS) plastic. Physical properties of ABS plastic related to radiation interaction and dosimetry were compared with commonly available phantom materials for dose measurements in radiotherapy. The ABS IMRT phantom has provisions to hold various types of detectors such as ion chambers, radiographic/radiochromic films, TLDs, MOSFETs, and gel dosimeters. The measurements related to pretreatment dose verification in IMRT of carcinoma prostate were carried out using ABS and Scanditronix-Wellhofer RW3 IMRT phantoms for five different cases. Point dose data were acquired using ionization chamber and TLD discs, while Gafchromic EBT and radiographic EDR2 films were used for generating 2D dose distributions. Treatment planning system (TPS) calculated and measured doses in ABS plastic and RW3 IMRT phantom were in agreement within . The dose values at a point in a given patient acquired using ABS and RW3 phantoms were found comparable within 1%. Fluence maps and dose distributions of these patients generated by TPS and measured in ABS IMRT phantom were also found comparable both numerically and spatially. This study indicates that ABS plastic IMRT phantom is a tissue-equivalent phantom and, dosimetrically, it is similar to solid/plastic water IMRT phantoms. Although this material is demonstrated for IMRT dose verification, it can also be used as a tissue-equivalent phantom material for other dosimetry purposes in radiotherapy. PACS number: 87.53Kn, 87.55Qr, 87.53Bn and 87.55Km
Keywords
Funding Information
- International Atomic Energy Agency
This publication has 13 references indexed in Scilit:
- Dosimetric pre-treatment verification of IMRT using an EPID; clinical experienceRadiotherapy and Oncology, 2006
- An inter-centre quality assurance network for IMRT verification: Results of the ESTRO QUASIMODO projectRadiotherapy and Oncology, 2005
- Dose verification in clinical imrt prostate incidentsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in MedicineInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- IMRT: delivery techniques and quality assuranceBritish Journal of Radiology, 2003
- Intensity-modulated radiotherapy of head-and-neck cancer: encouraging early resultsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- The effect of setup uncertainty on normal tissue sparing with IMRT for head-and-neck cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2001
- Intensity-Modulated Radiation Therapy2001
- Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracyMedical Physics, 2000
- Clinical experience with intensity modulated radiation therapy (IMRT) in prostate cancerRadiotherapy and Oncology, 2000