A new safe laparoscopic manipulator system with a five-bar linkage mechanism and an optical zoom
Open Access
- 1 January 1999
- journal article
- research article
- Published by Taylor & Francis in Computer Aided Surgery
- Vol. 4 (4) , 182-192
Abstract
This paper describes a new type of laparoscopic manipulator system designed with regard to its safety and sterilization in medical settings. It consists of a laparoscope manipulator using a five-bar linkage mechanism, an optical zoom called the automatic microzoom laparoscope, and a man-machine interface called the Head-Mouse system. The movement of the laparoscope manipulator in the X-Y plane determines the angle of the laparoscope, and the optical zoom substitutes for back-and-forth movements. This system achieves intrinsic safety and ease of sterilization due to limitations in the range of movement. The system achieves high reliability because of the simplicity of both its mechanism and its software. The newly developed optical zoom avoids the possibility of contact with organs by eliminating the back-and-forth movements of the laparoscope required for close-up observation while preserving a high-quality laparoscopic image. The Head-Mouse system for man-machine interface was designed to avoid incorrect...Keywords
This publication has 5 references indexed in Scilit:
- The evolutionary design of an endoscopic telemanipulatorIEEE Robotics & Automation Magazine, 1996
- Advanced Techniques in Gasless Laparoscopic Surgery1995
- The Internet and the World Wide Web:A General Overview and Applications for the Biomedical CommunityComputer Aided Surgery, 1995
- Controlling the movement of a surgical laparoscopeIEEE Engineering in Medicine and Biology Magazine, 1995
- The Development of a Surgeon Robot for ProstatectomiesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1991