Second-generation Highly Cross-linked X3™ Polyethylene Wear: A Preliminary Radiostereometric Analysis Study
- 1 October 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 468 (10) , 2704-2709
Abstract
Eration highly cross-linked polyethylene, which has improved wear and mechanical properties and resistance to oxidation in vitro. Owing to its recent introduction, there are no publications describing its clinical performance. Questions/purposes We assessed early clinical wear of a second-generation highly cross-linked polyethylene liner and compared its clinical performance with the published results of hip simulator tests and with first-generation highly cross-linked polyethylene annealed liners. Patients and Methods Twenty-one patients were enrolled in a prospective cohort study. Clinical outcome and femoral head penetration were measured for 19 patients at 6 months and 1 and 2 years postoperatively. Results The median proximal head penetration was 0.009 mm and 0.024 mm at 1 and 2 years, respectively. The median two-dimensional (2-D) head penetration was 0.083 mm and 0.060 mm at 1 and 2 years, respectively. The median proximal wear rate between 1 and 2 years was 0.015 mm/year. Conclusions The wear rate calculated was similar to the in vitro wear rate reported for this material; however, it was less than the detection threshold for this technique. Although longer followup is required for wear to reach a clinically quantifiable level, this low level of wear is encouraging for the future clinical performance of this material. Level of Evidence Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence....Keywords
This publication has 52 references indexed in Scilit:
- Wear of a highly cross-linked polyethylene liner: a preliminary RSA studyEuropean Journal of Orthopaedic Surgery & Traumatology, 2009
- Notes on the use and interpretation of radiostereometric analysisActa Orthopaedica, 2009
- The creep and wear of highly cross-linked polyethyleneThe Journal of Bone and Joint Surgery. British volume, 2008
- Standing versus Supine Radiographs in RSA Evaluation of Femoral Head PenetrationClinical Orthopaedics and Related Research, 2006
- RSA wear measurements with or without markers in total hip arthroplastyJournal of Biomechanics, 2006
- Weight bearing radiographs are not necessary for measurement of polyethylene penetration in total hip prostheses: A radiostereometric study of 111 patients examined in weight-bearing and supine positionActa Orthopaedica, 2006
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- OsteolysisClinical Orthopaedics and Related Research, 2004
- Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral headThe Journal of Bone and Joint Surgery. British volume, 2002
- Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene: A survey of 8 reports in the literatureActa Orthopaedica, 2001