This paper describes a method for increasing the accuracy and precision of temperature measurements of a liquid based on the central limit theorem. A thermometer immersed in a liquid exhibits a response with determined accuracy and precision. This measurement is integrated with an instrumentation and control system that imposes the behavioral conditions of the central limit theorem (CLT). The oversampling method exhibited an increasing measurement resolution. Through periodic sampling of large groups, an increase in the accuracy and formula of the increase in precision is developed. A measurement group sequencing algorithm and experimental system were developed to obtain the results of this system. Hundreds of thousands of experimental results are obtained and seem to demonstrate the proposed idea’s validity.Keywords
Funding Information
- Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
- Fundação de Apoio à Pesquisa do Estado da Paraíba (FAPESQ)
This publication has 11 references indexed in Scilit:
- Analysis of Sensors Noise Performance Using Allan Deviation2019
- Accurate Microwave Measurement of Coating Thickness on Carbon Composite SubstratesIEEE Transactions on Instrumentation and Measurement, 2016
- A theoretical study on the accuracy of the T-history method for enthalpy–temperature curve measurement: analysis of the influence of thermal gradients inside T-history samplesMeasurement Science and Technology, 2015
- Aggregating for accuracy: a closer look at sampling and accuracy in class action litigationLaw, Probability and Risk, 2014
- WHEN BIAS KILLS THE VARIANCE: CENTRAL LIMIT THEOREMS FOR DEA AND FDH EFFICIENCY SCORESEconometric Theory, 2014
- Revisiting Central Limit Theorem: Accurate Gaussian Random Number Generation in VLSIIEEE Transactions on Very Large Scale Integration (VLSI) Systems, 2014
- Measurement and statistical analysis toward reproducibility validation of AZ4562 cylindrical microlenses obtained by reflowMeasurement, 2014
- The International Temperature Scale of 1990 (ITS-90)Metrologia, 1990
- Should the classical variance be used as a basic measure in standards metrology?IEEE Transactions on Instrumentation and Measurement, 1987