Development and Validation of a Novel Calibration Methodology and Control Approach for Robot-Aided Transcranial Magnetic Stimulation (TMS)
- 29 January 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 68 (5) , 1589-1600
Abstract
Objective: This article presents the development and validation of a new robotic system for Transcranial Magnetic Stimulation (TMS), characterized by a new control approach, and an ad-hoc calibration methodology, specifically devised for the TMS application. Methods: The robotic TMS platform is composed of a 7 dof manipulator, controlled by an impedance control, and a camera-based neuronavigation system. The proposed calibration method was optimized on the workspace useful for the specific TMS application (spherical shell around the subject's head), and tested on three different hand-eye and robot-world calibration algorithms. The platform functionality was tested on six healthy subjects during a real TMS procedure, over the left primary motor cortex. Results: employing our method significantly decreases ( $p< 0.001$ ) the calibration error by 34% for the position and 19% for the orientation. The robotic TMS platform achieved greater orientation accuracy than the expert operators, significantly reducing orientation errors by 46% ( $p< 0.001$ ). No significant differences were found in the position errors and in the amplitude of the motor evoked potentials (MEPs) between the robot-aided TMS and the expert operators. Conclusion: The proposed calibration represents a valid method to significantly reduce the calibration errors in robot-aided TMS applications. Results showed the efficacy of the proposed platform (including the control algorithm) in administering a real TMS procedure, achieving better coil positioning than expert operators, and similar results in terms of MEPs. Significance: This article spotlights how to improve the performance of a robotic TMS platform, providing a reproducible and low-cost alternative to the few devices commercially available.Keywords
Funding Information
- European Research Council
- Restoring the Self with Embodiable Hand Prostheses (678908)
This publication has 46 references indexed in Scilit:
- Optimal Coil Orientation for Transcranial Magnetic StimulationPLOS ONE, 2013
- Non‐orthogonal tool/flange and robot/world calibrationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2012
- Robust real‐time robot–world calibration for robotized transcranial magnetic stimulationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2011
- Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and researchClinical Neurophysiology, 2009
- Robot-assisted image-guided transcranial magnetic stimulation for somatotopic mapping of the motor cortex: a clinical pilot studyActa Neurochirurgica, 2009
- Optimizing Functional Accuracy of TMS in Cognitive Studies: A Comparison of MethodsJournal of Cognitive Neuroscience, 2009
- Transcranial magnetic stimulation and the challenge of coil placement: A comparison of conventional and stereotaxic neuronavigational strategiesHuman Brain Mapping, 2007
- Modeling and copying human head movementsIEEE Transactions on Robotics and Automation, 1999
- Force regulation and contact transition controlIEEE Control Systems, 1996