Video Methods for Evaluating Physiologic Monitor Alarms And Alarm Responses
Authors:
- Christopher P. BonafideCBChristopher P. Bonafide
- Miriam ZanderMZMiriam Zander
- Christian Sarkis GrahamCGChristian Sarkis Graham
- Christine M. Weirich PaineCPChristine M. Weirich Paine
- Whitney RockWRWhitney Rock
- Andrew RichARAndrew Rich
- Kathryn E. RobertsKRKathryn E. Roberts
- Margaret FortinoMFMargaret Fortino
- Vinay M. NadkarniVNVinay M. Nadkarni
- Richard LinRLRichard Lin
- Ron KerenRKRon Keren
- 20 May 2014
- journal article
- Published by Association for the Advancement of Medical Instrumentation (AAMI) in Biomedical Instrumentation & Technology
- Vol. 48 (3) , 220-230
Abstract
False physiologic monitor alarms are extremely common in the hospital environment. High false alarm rates have the potential to lead to alarm fatigue, leading nurses to delay their responses to alarms, ignore alarms, or disable them entirely. Recent evidence from the U.S. Food and Drug Administration (FDA) and The Joint Commission has demonstrated a link between alarm fatigue and patient deaths. Yet, very little scientific effort has focused on the rigorous quantitative measurement of alarms and responses in the hospital setting. We developed a system using multiple temporarily mounted, minimally obtrusive video cameras in hospitalized patients' rooms to characterize physiologic monitor alarms and nurse responses as a proxy for alarm fatigue. This allowed us to efficiently categorize each alarm's cause, technical validity, actionable characteristics, and determine the nurse's response time. We describe and illustrate the methods we used to acquire the video, synchronize and process the video, manage the large digital files, integrate the video with data from the physiologic monitor alarm network, archive the video to secure servers, and perform expert review and annotation using alarm "bookmarks." We discuss the technical and logistical challenges we encountered, including the root causes of hardware failures as well as issues with consent, confidentiality, protection of the video from litigation, and Hawthorne-like effects. The description of this video method may be useful to multidisciplinary teams interested in evaluating physiologic monitor alarms and alarm responses to better characterize alarm fatigue and other patient safety issues in clinical settings.Keywords
This publication has 20 references indexed in Scilit:
- Validation of the Work Observation Method By Activity Timing (WOMBAT) method of conducting time-motion observations in critical care settings: an observational studyBMC Medical Informatics and Decision Making, 2011
- Intensive care unit alarms—How many do we need?*Critical Care Medicine, 2010
- Monitor Alarm Fatigue: Standardizing Use of Physiological Monitors and Decreasing Nuisance AlarmsAmerican Journal of Critical Care, 2010
- Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical ContextAnesthesia & Analgesia, 2009
- Research electronic data capture (REDCap)—A metadata-driven methodology and workflow process for providing translational research informatics supportJournal of Biomedical Informatics, 2009
- Video capture of clinical care to enhance patient safetyQuality and Safety in Health Care, 2004
- Human probability matching behaviour in response to alarms of varying reliabilityErgonomics, 1995
- System operator response to warnings of danger: A laboratory investigation of the effects of the predictive value of a warning on human response time.Journal of Experimental Psychology: Applied, 1995
- Crying wolfCritical Care Medicine, 1994