Abstract
Purpose: To determine the incidence of keratitis and conjunctivitis associated with reusable contact lens use during ophthalmic laser procedures at an academic medical center. Patients and Methods: All laser procedures were performed at a single nonprofit academic medical center (Jacksonville, FL, USA). Data from patients that underwent neodymium-doped yttrium-aluminum-garnet (Nd:YAG) capsulotomy or selective laser trabeculoplasty (SLT) with reusable contact lenses between 2014 and 2023 was analyzed retrospectively. The post-procedure infection rate, defined as the incidence of newly diagnosed superficial keratitis and/or conjunctivitis within two weeks of the laser procedure, was calculated by dividing the total number of affected eyes by the total cohort size. A break-even analysis was conducted to evaluate the economic impact of reusable lens implementation. Results: A total of 2285 eyes of 1363 patients were included in this study. The incidence of previous conjunctivitis in this cohort was small (0.4%) and no eyes had a prior history of keratitis. Of the 1372 Nd:YAG capsulotomy and 913 SLT procedures performed, only one eye (0.044%) developed signs of post-operative conjunctivitis and none developed keratitis. The single case of unilateral conjunctivitis occurred within one day of a Nd:YAG capsulotomy and was secondary to medicamentosa. A cost analysis showed that reusable ophthalmic laser lenses become more cost-effective than disposable lenses after approximately 86 SLT procedures and 91 Nd:YAG capsulotomies, while reducing medical waste by 79%. Conclusion: Given the negligible rate of conjunctivitis, substantial cost savings, and significant reduction in environmental waste associated with reusable capsulotomy and trabeculoplasty lenses, the routine use of disposable ophthalmic lenses should be reconsidered.Keywords
This publication has 32 references indexed in Scilit:
- A Comparative Analysis of Intraocular Pressure Measurement Accuracy With Reused iCare ProbesOphthalmology Glaucoma, 2024
- Dangerous traps: Macroplastic encounters affecting freshwater and terrestrial wildlifeScience of The Total Environment, 2021
- Microplastics in Seafood and the Implications for Human HealthCurrent Environmental Health Reports, 2018
- Disinfection of TonometersOphthalmology, 2017
- Use of Various Glaucoma Surgeries and Procedures in Medicare Beneficiaries from 1994 to 2012Ophthalmology, 2015
- Disinfection, Sterilization, and Control of Hospital Waste2015
- New developments in reprocessing semicritical itemsAmerican Journal of Infection Control, 2013
- Contamination of disposable tonometer prisms during tonometryEye, 2005
- Use of Nd:YAG laser capsulotomySurvey of Ophthalmology, 2003
- Probability of Nd:YAG laser capsulotomy after cataract surgery in Olmsted County, MinnesotaAmerican Journal of Ophthalmology, 2001