Comparison of two electronic apex locators on human cadavers
- 4 November 2015
- journal article
- research article
- Published by Springer Nature in Clinical Oral Investigations
- Vol. 20 (7) , 1547-1550
Abstract
The aim of this study on human cadavers was to compare the accuracy of two electronic apex locators (EALs) Dentaport ZX (J. Morita Corporation, Tokyo, Japan) and Raypex 6 (VDW, Munich, Germany). Twenty-two single rooted teeth of four human cadaver heads were scheduled for this study. Before the extraction, an access cavity was opened and the crown was cut to establish a stable reference point for all measurements. The working length determination was performed with Dentaport ZX and Raypex 6 in the presence or not of 5.25 % sodium hypochlorite (SH) using a k-file 10. The teeth were then extracted and the real working length (RWL) was measured under a stereomicroscope at ×30 magnification. The difference between the two working lengths was calculated: positive values indicate measurements exceeding the foramen, while negative values indicated measurements short of the foramen. The data were analyzed with a t test analysis. The mean of distances was 0.33 ± 0.20 mm and 0.32 ± 0.2 mm for Dentaport ZX respectively in the presence or not of SH and 0.38 ± 0.20 mm and 0.39 ± 0.19 mm for Raypex 6. No statistical differences were found between the two devices (p > 0.05). Both apex locators showed a high accuracy in the presence or not of SH. Both electronic apex locators can be recommended for clinical use and their accuracy is not affected by SH.Keywords
This publication has 30 references indexed in Scilit:
- Accuracy of working length determination using 3 electronic apex locators and direct digital radiographyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2011
- Comparison of working length determination with radiographs and two electronic apex locatorsInternational Endodontic Journal, 2009
- Power analysis and sample size calculationsCurrent Anaesthesia & Critical Care, 2008
- A comparison between two electronic apex locators: an ex vivo investigationInternational Endodontic Journal, 2007
- Evaluation of working length determination methods: An in vivo / ex vivo studyIndian Journal of Dental Research, 2007
- Antimicrobial efficacy of ozonated water, gaseous ozone, sodium hypochlorite and chlorhexidine in infected human root canalsInternational Endodontic Journal, 2006
- Electronic apex locatorsInternational Endodontic Journal, 2004
- Success rate of endodontic treatment of teeth with vital and nonvital pulps. A meta-analysisOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2004
- Electronic apex locatorDental Clinics of North America, 2004
- The position and topography of the apical canal constriction and apical foramenInternational Endodontic Journal, 1984