Discrepancies between isotope ratio infrared spectroscopy and isotope ratio mass spectrometry for the stable isotope analysis of plant and soil waters
- 16 June 2010
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 24 (14) , 1948-1954
Abstract
The use of isotope ratio infrared spectroscopy (IRIS) for the stable hydrogen and oxygen isotope analysis of water is increasing. While IRIS has many advantages over traditional isotope ratio mass spectrometry (IRMS), it may also be prone to errors that do not impact upon IRMS analyses. Of particular concern is the potential for contaminants in the water sample to interfere with the spectroscopy, thus leading to erroneous stable isotope data. Water extracted from plant and soil samples may often contain organic contaminants. The extent to which contaminants may interfere with IRIS and thus impact upon data quality is presently unknown. We tested the performance of IRIS relative to IRMS for water extracted from 11 plant species and one organic soil horizon. IRIS deviated considerably from IRMS for over half of the samples tested, with deviations as large as 46‰ (δ 2 H) and 15.4‰ (δ 18 O) being measured. This effect was reduced somewhat by using activated charcoal to remove organics from the water; however, deviations as large as 35‰ (δ 2 H) and 11.8‰ (δ 18 O) were still measured for these cleaned samples. Interestingly, the use of activated charcoal to clean water samples had less effect than previously thought for IRMS analyses. Our data show that extreme caution is required when using IRIS to analyse water samples that may contain organic contaminants. We suggest that the development of new cleaning techniques for removing organic contaminants together with instrument‐based software to flag potentially problematic samples are necessary to ensure accurate plant and soil water analyses using IRIS. Copyright © 2010 John Wiley & Sons, Ltd.Keywords
This publication has 25 references indexed in Scilit:
- High‐frequency field‐deployable isotope analyzer for hydrological applicationsWater Resources Research, 2009
- Demonstration of high‐precision continuous measurements of water vapor isotopologues in laboratory and remote field deployments using wavelength‐scanned cavity ring‐down spectroscopy (WS‐CRDS) technologyRapid Communications in Mass Spectrometry, 2009
- Cavity ring‐down spectroscopy versus high‐temperature conversion isotope ratio mass spectrometry; a case study on δ 2 H and δ 18 O of pure water samples and alcohol/water mixturesRapid Communications in Mass Spectrometry, 2009
- High-Precision Laser Spectroscopy D/H and 18O/16O Measurements of Microliter Natural Water SamplesAnalytical Chemistry, 2007
- Seasonal variations in moisture use in a piñon–juniper woodlandOecologia, 2007
- Toward a plant‐based proxy for the isotope ratio of atmospheric water vaporGlobal Change Biology, 2007
- Continuous flow 2 H/ 1 H and 18 O/ 16 O analysis of water samples with dual inlet precisionRapid Communications in Mass Spectrometry, 2004
- New Methods for Fully Automated Isotope Ratio Determination from Hydrogen at the Natural Abundance LevelIsotopes in Environmental and Health Studies, 1996
- Sample Preparation Device for Quantitative Hydrogen Isotope Analysis Using Chromium MetalAnalytical Chemistry, 1996
- Extraction technique for the determination of oxygen-18 in water using preevacuated glass vialsAnalytical Chemistry, 1992