Component testing for x-ray spectroscopy and polarimetry
Authors:
- 9 September 2019
- conference paper
- Published by SPIE-Intl Soc Optical Eng
- Vol. 11118, 36
Abstract
We present the performance and recent results of the MIT polarimetry beamline. Originally designed for testing Chandra HETG gratings, the beamline has been adapted to test components for soft x-ray polarimetry applications. Since then, its monochromator capabilities have also been used to test gratings. We present results on the measured absolute efficiencies of the Arcus Phase A gratings using the B-K, O-K, and C-K emission lines. The beamline has also been used to develop tools and techniques to measure the linear polarization of soft X-rays (0.2-0.8 keV), which form the basis for a sounding rocket mission REDSoX (Rocket Experiment Demonstration of a Soft X-ray Polarimeter) and a possible orbital mission. We present our tests to align the REDSoX gratings, as well as our idea to use thin twisted crystals as a possible alternative to laterally-graded multilayer mirrors. Support for this work was provided in part by the NASA grant NNX15AL14G and a grant from the MIT Kavli Institute.Keywords
This publication has 1 reference indexed in Scilit:
- Design of a broadband soft x-ray polarimeterJournal of Astronomical Telescopes, Instruments, and Systems, 2018