Meta-analytic evaluation of complications and longevity of bone-anchored prostheses in lower extremity amputees: does design matter?
Authors:
- Hugh MilchemHMHugh Milchem
- Yasmine J. KhairYKYasmine J. Khair
- Maamoun AdraMAMaamoun Adra
- Shayndhan S. SivanathanSSShayndhan S. Sivanathan
- Shreehari SureshSSShreehari Suresh
- Dilip K. VankayalapatiDVDilip K. Vankayalapati
- Aslam Mohamed HaroonAHAslam Mohamed Haroon
- Christian A. ThanCTChristian A. Than
- Nadim TaraziNTNadim Tarazi
- Hayato NakanishiHNHayato Nakanishi
- Peter J. SmithamPSPeter J. Smitham
Open Access
- 14 January 2026
- journal article
- review article
- Published by Springer Nature in MUSCULOSKELETAL SURGERY
- Vol. 110 (3) , 1-15
Abstract
Bone-anchored prostheses (BAPs) are an alternative option for lower-limb amputees with problematic suspended socket prostheses (SSPs). We sought to meta-analytically quantify complication burden and revision-free survival for BAPs, whilst investigating possible differences in complication rates between common screw-fit (OPRA) and press-fit (ILP/OPL) designs. A multi-database search of PubMed, EMBASE, CiNAHL, Cochrane Cochrane Central Register of Controlled Trials, Cochrane Database of Systematic Reviews, Scopus and Web of Science from inception to September 2025 was conducted. Proportions were pooled using a random-effects model. This review was registered in PROPSPERO (ID: CRD42024507070). A total of 22 studies reporting 979 patients were included. An overall complication rate of 65% (95% CI: 0.53–0.75, I 2 = 69%) was found. The most common complication was breakage of the abutment or dual-cone adaptor (DCA) (0.38, 95% CI: 0.19–0.62, I 2 = 97%), which constituted 94% of all mechanical failures (95% CI: 0.87–0.98, I 2 = 51%). Second was infection (0.36, 95% CI: 0.25–0.50, I 2 = 92%) with 72% of infections being grade I (95% CI: 0.57–0.83, I 2 = 72%). Estimated revision-free survival was 89% (95% CI: 0.84–0.93, I 2 = 0%) at five years and 77% (95% CI: 0.70–0.82, I 2 = 0%) at ten years. The largest difference between implant designs was for overall proportion of mechanical failure, OPRA at 68% (95% CI: 0.34–0.90, I 2 = 98%), ILP/OPL at 19% (0.09–0.36, I 2 = 85%). BAPs do carry appreciable risks for superficial infection and mechanical (abutment/DCA) breakage, with screw-fit designs appearing to have a higher mechanical failure rate. However, the revision-free survival rate overall remains acceptable up to 10 years demonstrating the viability of BAPs in lower-limb amputees.Keywords
This publication has 52 references indexed in Scilit:
- Endo-Exo-Prothesen nach GliedmaßenamputationDer Orthopäde, 2015
- Fifteen years of experience with Integral-Leg-Prosthesis: Cohort study of artificial limb attachment systemJournal of Rehabilitation Research and Development, 2015
- Predictors of non-use of prostheses by people with lower limb amputation after discharge from rehabilitation: development and validation of clinical prediction rulesJournal of Physiotherapy, 2014
- Outcome of Percutaneous Osseointegrated Prostheses for Patients With Unilateral Transfemoral Amputation at Two-Year Follow-UpArchives of Physical Medicine and Rehabilitation, 2014
- Application and survival curve of total hip arthroplasties: a systematic comparative analysis using worldwide hip arthroplasty registersInternational Orthopaedics, 2012
- Enhanced secondary analysis of survival data: reconstructing the data from published Kaplan-Meier survival curvesBMC Medical Research Methodology, 2012
- Skin Problems of the Stump in Lower Limb Amputees: 1. A Clinical StudyActa Dermato-Venereologica, 2011
- Advances in Lower-limb Prosthetic TechnologyPhysical Medicine and Rehabilitation Clinics of North America, 2010
- Determinants of Skin Problems of the Stump in Lower-Limb AmputeesArchives of Physical Medicine and Rehabilitation, 2009
- Socket Versus Bone-Anchored Trans-Femoral ProsthesesProsthetics and Orthotics International, 2005