Deep learning-based 3D in vivo dose reconstruction with an electronic portal imaging device for magnetic resonance-linear accelerators: a proof of concept study
- 2 December 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (23) , 235011
Abstract
Objective. To develop a novel deep learning-based 3D in vivo dose reconstruction framework with an electronic portal imaging device (EPID) for magnetic resonance-linear accelerators (MR-LINACs). Approach. The proposed method directly back-projected 2D portal dose into 3D patient coarse dose, which bypassed the complicated patient-to-EPID scatter estimation step used in conventional methods. A pre-trained convolutional neural network (CNN) was then employed to map the coarse dose to the final accurate dose. The electron return effect caused by the magnetic field was captured with the CNN model. Patient dose and portal dose datasets were synchronously generated with Monte Carlo simulation for 96 patients (78 cases for training and validation and 18 cases for testing) treated with fixed-beam intensity-modulated radiotherapy in four different tumor sites, including the brain, nasopharynx, lung, and rectum. Beam angles from the training dataset were further rotated 2–3 times, and doses were recalculated to augment the datasets. Results. The comparison between reconstructed doses and MC ground truth doses showed mean absolute errors γ -passing rates (3%, 2 mm) were 97.42%±2.66% (brain), 98.53%±0.95% (nasopharynx), 99.41%±0.46% (lung), and 98.63%±1.01% (rectum). The dose volume histograms and indices also showed good consistency. The average dose reconstruction time, including back projection and CNN dose mapping, was less than 3 s for each individual beam. Significance. The proposed method can be potentially used for accurate and fast 3D dosimetric verification for online adaptive radiotherapy using MR-LINACs.Keywords
Funding Information
- Guangdong Basic and Applied Basic Research Foundation, China (2021A1515012044)
- Guangzhou Science and Technology Foundation, China (202102020968)
- National Natural Science Foundation of China (11805292)
This publication has 42 references indexed in Scilit:
- Integrated megavoltage portal imaging with a 1.5 T MRI linacPhysics in Medicine & Biology, 2011
- 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetryRadiotherapy and Oncology, 2010
- Direct-detection EPID dosimetry: investigation of a potential clinical configuration for IMRT verificationPhysics in Medicine & Biology, 2009
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strengthPhysics in Medicine & Biology, 2008
- Routine individualised patient dosimetry using electronic portal imaging devicesRadiotherapy and Oncology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- Dosimetric characteristics of a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995