Developing a device model for a safe and effective method in catheter fixation: An experimental study
- 9 March 2026
- journal article
- Published by Ege Journal of Medicine in Ege Tıp Dergisi
- Vol. 65 (1) , 49-54
Abstract
Aim: Current permanent catheter fixation techniques may lead to epidural catheter dislodgement or extrusion, resulting in risks such as catheter migration, inadequate analgesia, and unintended spinal block. This study aimed to compare a newly developed fixation device with the conventional suture method used for epidural catheter fixation.Materials and Methods: The Epilock® device, developed at Ege University Technopark CND ARGE, was manufactured using Polylactic Acid Thermoplastic Polymer (PLA+) material via a 3D printer. After obtaining ethical approval, 10 Sprague Dawley rats (5 males, 5 females) were anesthetized and randomly allocated into two groups. In the Control Group (n=5), epidural catheters were fixed 11 cm beneath the skin with silk sutures (20 knots). In the Experimental Group (n=5), catheters were secured using the Epilock® device. Fixation time was recorded, and the rats were monitored for catheter stability and infection for 7 days.Results: In the Control Group (n=5), the mean catheter fixation time was 258 ±250 seconds, while in the Experimental Group (n=5) it was 175 ±150 seconds. The Epilock® device significantly reduced fixation time compared the Control Group (p=0.032). Catheters remained significantly longer in place in the Experimental Group on the first day (p=0.048), whereas no significant difference was found on the second day (p=1.000).Conclusion: The Epilock® device provided a shorter procedure time and improved initial catheter retention compared with conventional suturing. These findings suggest it may enhance patient comfort, especially for those with positioning difficulties, and has the potential to be a valuable option in routine practice with further development and larger studies.Keywords
Funding Information
- This experimental study was supported by CND R&D, operating within Ege University Technopark
This publication has 16 references indexed in Scilit:
- Comparative efficacy of Micropore™ surgical dressing, Tegaderm™ and Lockit plus® for lumbar epidural catheter fixation in children: a prospective parallel group randomized controlled trialRevista Española de Anestesiología y Reanimación (English Edition), 2023
- Epidural analgesia for postoperative pain: Improving outcomes or adding risks?Best Practice & Research Clinical Anaesthesiology, 2021
- Complications of Postoperative Epidural Analgesia For Oncologic SurgeryThe Clinical Journal of Pain, 2019
- Tunnelling of thoracic epidural catheters is associated with fewer catheter-related infections: a retrospective registry analysisBritish Journal of Anaesthesia, 2016
- Nationwide incidence of serious complications of epidural analgesia in the United StatesActa Anaesthesiologica Scandinavica, 2016
- Failed epidural: causes and managementBritish Journal of Anaesthesia, 2012
- The effect of the Lockit epidural catheter clamp on epidural migration: a controlled trialAnaesthesia, 2001
- Postoperative pain relief using thoracic epidural analgesia: outstanding success and disappointing failuresAnaesthesia, 2001
- Epidural catheter fixation: subcutaneous tunnelling with a loop to prevent displacementAnaesthesia, 2000
- Securing epidural cathetersCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1982