A short tutorial contribution to impedance and AC-electrokinetic characterization and manipulation of cells and media: Are electric methods more versatile than acoustic and laser methods?
Open Access
- 1 January 2014
- journal article
- Published by Walter de Gruyter GmbH in Journal of Electrical Bioimpedance
- Vol. 5 (1) , 74-91
Abstract
Lab-on-chip systems (LOCs) can be used as in vitro systems for cell culture or manipulation in order to analyze or monitor physiological cell parameters. LOCs may combine microfluidic structures with integrated elements such as piezo-transducers, optical tweezers or electrodes for AC-electrokinetic cell and media manipulations. The wide frequency band (1 GHz) usable for AC-electrokinetic manipulation and characterization permits avoiding electrochemical electrode processes, undesired cell damage, and provides a choice between different polarization effects that permit a high electric contrast between the cells and the external medium as well as the differentiation between cellular subpopulations according to a variety of parameters. It has been shown that structural polarization effects do not only determine the impedance of cell suspensions and the force effects in AC-electrokinetics but can also be used for the manipulation of media with inhomogeneous temperature distributions. This manuscript considers the interrelations of the impedance of suspensions of cells and AC-electrokinetic single cell effects, such as electroorientation, electrodeformation, dielectrophoresis, electrorotation, and travelling wave (TW) dielectrophoresis. Unified models have allowed us to derive new characteristic equations for the impedance of a suspension of spherical cells, TW dielectrophoresis, and TW pumping. A critical review of the working principles of electro-osmotic, TW and electrothermal micropumps shows the superiority of the electrothermal pumps. Finally, examples are shown for LOC elements that can be produced as metallic structures on glass chips, which may form the bottom plate for self-sealing microfluidic systems. The structures can be used for cell characterization and manipulation but also to realize micropumps or sensors for pH, metabolites, cell-adhesion, etc.Keywords
This publication has 73 references indexed in Scilit:
- Acoustofluidics 23: acoustic manipulation combined with other force fieldsLab on a Chip, 2013
- Experimental verification of an equivalent circuit for the characterization of electrothermal micropumps: High pumping velocities induced by the external inductance at driving voltages below 5 VElectrophoresis, 2012
- Micro Total Analysis Systems for Cell Biology and Biochemical AssaysAnalytical Chemistry, 2011
- Single-cell microfluidic impedance cytometry: a reviewMicrofluidics and Nanofluidics, 2010
- Review of cell and particle trapping in microfluidic systemsAnalytica Chimica Acta, 2009
- Traveling wave dielectrophoresis micropump based on the dispersion of a capacitive electrode layerJournal of Applied Physics, 2009
- On the field distribution in electrorotation chambers—Influence of electrode shapeElectrochimica Acta, 2006
- Alpha- and beta-dispersion of fixed platelets: comparison with a structure-based theoretical approachColloids and Surfaces A: Physicochemical and Engineering Aspects, 2002
- The use of neuronal networks on multielectrode arrays as biosensorsBiosensors and Bioelectronics, 1995
- Orientation of human and avian erythrocytes in radio-frequency fieldsExperimental Cell Research, 1970