IOL Calculators Variability in Toric Lens Planning: A Retrospective Paired Comparison of the Planning Discrepancies Between the Johnson & Johnson Toric Calculator and ESCRS-Hosted Engines
Open Access
- 1 June 2026
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 20, 1-13
Abstract
Purpose: To determine the frequency, magnitude, and direction of toric intraocular lens (IOL) planning discrepancies between the Johnson & Johnson (J&J) official calculator and European Society of Cataract and Refractive Surgeons (ESCRS)-hosted toric engines, and to identify preoperative predictors of clinically relevant disagreements. Methods: This retrospective, single-center study included 182 eyes of 127 patients undergoing toric IOL implantation (TECNIS). For each eye, planning results were generated using identical biometric data (IOLMaster 700, CASIA2 keratometry) across all calculators. The primary comparison was J&J with posterior corneal astigmatism correction enabled (PCA-ON) versus ESCRS-Barrett – secondary comparisons included Kane, EVO 2.0, and Hoffer QST. Clinically relevant discrepancy (ClinRel) was defined as a toric cylinder step difference ≥ 1 and/or axis difference ≥ 10°. Logistic regression, including Firth-penalized estimation, identified predictors of ClinRel. The primary analysis set comprised one eye per patient (n=127). Results: ClinRel between J&J PCA-ON and Barrett occurred in 59.8% of eyes (76/127; 95% CI: 51.1– 68.0%), driven almost only by step differences. J&J PCA-ON recommended a higher step in 56.7% of eyes, with only 2.4% showing the opposite. Discrepancies were concentrated in with-the-rule eyes (72.3%) and nearly absent in against-the-rule eyes (5.6%). Rates were higher for secondary engines (Kane 83.8%, EVO 74.4%, Hoffer QST 76.7%). An Alpins-style paired-planning analysis yielded a Correction Index of 1.20, with the Difference Vector concentrated at the vertical meridian (84°). Posterior corneal astigmatism grade was the strongest independent predictor (Firth-adjusted OR = 203.5; p=0.014). An empirical PCA threshold of 0.36 D (Youden index; sensitivity 91.6%, specificity 56.0%) identified eyes at elevated risk of calculator disagreement. Conclusion: Calculators’ variability in toric IOL planning is substantial (~60% of eyes) and concentrated in with-the-rule eyes with elevated posterior corneal astigmatism. In this subgroup, consulting multiple calculators is recommended; in against-the-rule eyes, calculator choice rarely alters the recommendation.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of the accuracy of twenty intraocular lens power calculation formulas in long eyesAsia-Pacific Journal of Ophthalmology, 2025
- Accuracy of 7 artificial intelligence-based intraocular lens power calculation formulas in medium-long eyes: 2-center studyCanadian Journal of Ophthalmology, 2025
- Toric Calculations2024
- Risk factors for refractive error after cataract surgery: Analysis of 282 811 cataract extractions reported to the European Registry of Quality Outcomes for cataract and refractive surgeryJournal of Cataract & Refractive Surgery, 2018
- New algorithm for toric intraocular lens power calculation considering the posterior corneal astigmatismJournal of Cataract & Refractive Surgery, 2018
- Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: Comparison of methodologiesJournal of Cataract & Refractive Surgery, 2016
- Correcting astigmatism with toric intraocular lenses: Effect of posterior corneal astigmatismJournal of Cataract & Refractive Surgery, 2013
- Toric intraocular lenses: Historical overview, patient selection, IOL calculation, surgical techniques, clinical outcomes, and complicationsJournal of Cataract & Refractive Surgery, 2013
- Contribution of posterior corneal astigmatism to total corneal astigmatismJournal of Cataract & Refractive Surgery, 2012
- A solution to the problem of separation in logistic regressionStatistics in Medicine, 2002