Glistenings in foldable intraocular lenses
- 1 July 2002
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 28 (7) , 1211-1216
Abstract
To evaluate the formation of glistenings in foldable intraocular lenses (IOLs) of various materials after cataract surgery. University Eye Clinic of Trieste, Trieste, Italy. In this prospective study, 273 patients were randomized to have implantation of 1 of 7 types of foldable IOLs in the capsular bag after phacoemulsification. The IOLs were the CeeOn Edge 911A (Pharmacia & Upjohn Co.), ACR6D (Corneal), AcrySof (Alcon), SI-40NB (AMO), Hydroview H60M (Storz), Sensar (AMO), and Stabibag (Ioltech). The presence of glistenings in the IOL optic was assessed by slitlamp evaluation 7, 30, 90, 180, 360, and 720 days after surgery. The glistenings were graded at each control. Glistenings were observed in all IOL groups. The percentage of patients with glistenings ranged from 40.0% to 67.5%, with the percentage increasing up to 90 days after surgery and then becoming stable in most groups. The AcrySof group had a continuous increase over time. The mean grade of glistenings increased up to 180 days after surgery and then became stable in most groups. The AcrySof and 911A groups had a continuous increase over time. Glistening formation was observed in 7 different foldable IOLs. The AcrySof group had a higher percentage and a greater density of glistenings.Keywords
This publication has 15 references indexed in Scilit:
- Glistening formation in an AcrySof lens initiated by spinodal decompositionof the polymer network bytemperature changeJournal of Cataract & Refractive Surgery, 2001
- Glistenings in the AcrySof intraocular lens: pilot studyJournal of Cataract & Refractive Surgery, 2001
- Effect of round-edged acrylic intraocular lenses on preventing posterior capsule opacificationJournal of Cataract & Refractive Surgery, 2001
- Vacuoles in the AcrysofTM Intraocular Lens as Factor of the Presence of Serum in Aqueous HumorOphthalmic Research, 2001
- Intraindividueller Vergleich von Intraokularlinsen aus hochrefraktivem Silikon (Allergan SI40NB) und hydrophobem Acrylat (Alcon Acrysof MA60BM)Der Ophthalmologe, 2000
- Preventing lens epithelial cell migration using intraocular lenses with sharp rectangular edgesJournal of Cataract & Refractive Surgery, 2000
- Two year results: sharp versus rounded optic edges on silicone lensesJournal of Cataract & Refractive Surgery, 2000
- Optical and atomic force microscopy of an explanted AcrySof intraocular lens with glisteningsJournal of Cataract & Refractive Surgery, 2000
- Surgical prevention of posterior capsule opacificationJournal of Cataract & Refractive Surgery, 2000
- Practice styles and preferences of ASCRS members—1995 surveyJournal of Cataract & Refractive Surgery, 1996