A comparison of polysomnography and the WatchPAT in the diagnosis of obstructive sleep apnea

Authors:

Abstract
Objective: Our goal was to validate the WatchPAT in the diagnosis of obstructive sleep apnea.Study Design: We conducted a prospective, blinded, nonrandomized clinical trial.Methods: Patients with suspected obstructive sleep apnea scheduled for an overnight level I polysomnogram were offered enrollment in a study to compare the WatchPAT (Itamar Ltd, Israel) device with polysomnography. Patients wore the WatchPAT device simultaneously while undergoing polysomnography during evaluation in the sleep lab.Results: Thirty‐seven patients participated in the study. They had a mean age of 50.1 years (range, 31–73 years) and mean body mass index of 34.6 kg/m2 (range, 21.2–46.8 kg/m2). There was high correlation between the polysomnogram and WatchPAT apnea‐hypopnea index (r = 0.9288; 95% confidence interval = 0.8579–0.9650, P < 0.0001). The lowest oxygen saturation also showed high correlation (r = 0.989; 95% confidence interval = 0.9773–0.9947, P < 0.0001). The overall polysomnogram and WatchPAT sleep times revealed a correlation of r = 0.5815 (P = 0.005).Conclusion: The WatchPAT showed a high correlation with the polysomnogram in apnea‐hypopnea index, lowest oxygen saturation, and sleep time.Significance: It's use as a reliable tool in the diagnosis of Obstructive Sleep Apnea.

This publication has 14 references indexed in Scilit: