Abstract
A 3-D super-resolution (SR) pipeline based on data from a row-column (RC) array is presented. The 3-MHz RC array contains 62 rows and 62 columns with a half wavelength pitch. A synthetic aperture (SA) pulse inversion sequence with 32 positive and 32 negative row emissions is used for acquiring volumetric data using the SARUS research ultrasound scanner. Data received on the 62 columns are beamformed on a GPU for a maximum volume rate of 156 Hz when the pulse repetition frequency is 10 kHz. Simulated and 3-D printed point and flow microphantoms are used for investigating the approach. The flow microphantom contains a 100- [Formula: see text] radius tube injected with the contrast agent SonoVue. The 3-D processing pipeline uses the volumetric envelope data to find the bubble's positions from their interpolated maximum signal and yields a high resolution in all three coordinates. For the point microphantom, the standard deviation on the position is (20.7, 19.8, 9.1) [Formula: see text]. The precision estimated for the flow phantom is below [Formula: see text] in all three coordinates, making it possible to locate structures on the order of a capillary in all three dimensions. The RC imaging sequence's point spread function has a size of 0.58 × 1.05 × 0.31 mm3 ( 1.17λ×2.12λ×0.63λ ), so the possible volume resolution is 28900 times smaller than for SA RC B-mode imaging.Keywords
Funding Information
- Højteknologifonden (82-2014-4)
- Innovationsfonden (7050-00004B)
- BK Medical, Herlev, Denmark
This publication has 42 references indexed in Scilit:
- 3-D imaging using row-column-addressed arrays with integrated apodization - part i: apodization design and line element beamformingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2015
- 3-D imaging using row–column-addressed arrays with integrated apodization— part ii: transducer fabrication and experimental resultsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2015
- In Vivo Acoustic Super-Resolution and Super-Resolved Velocity Mapping Using MicrobubblesIEEE Transactions on Medical Imaging, 2014
- 2-D tissue motion compensation of synthetic transmit aperture imagesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- Top-orthogonal-to-bottom-electrode (TOBE) CMUT arrays for 3-D ultrasound imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- Beamforming of sound from two-dimensional arrays using spatial matched filtersThe Journal of the Acoustical Society of America, 2013
- A super‐resolution ultrasound method for brain vascular mappingMedical Physics, 2013
- Acoustic super-resolution with ultrasound and microbubblesPhysics in Medicine & Biology, 2013
- SARUS: A synthetic aperture real-time ultrasound systemIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2013
- Optimising phase and amplitude modulation schemes for imaging microbubble contrast agents at low acoustic powerUltrasound in Medicine & Biology, 2005