Thermal illuminators for far-infrared and submillimeter astronomical instruments
- 1 June 2005
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 44 (16) , 3208-3217
Abstract
Astronomical instruments operating in the infrared–millimeter region often require internal sources for detector monitoring, instrument calibration, and health checking. We describe the design, modeling, and experimental evaluation of thermal emitters with a fast speed of response and low-power dissipation, suitable for the far-infrared–submillimeter region. The development of an internal illuminator for the Herschel Spectral and Photometric Imaging Receiver (SPIRE) satellite instrument is used as an example of the optimization of the design to meet particular requirements. A prototype illuminator design for SPIRE has been developed that, for a power dissipation of 1.5 mW, provides an equivalent blackbody temperature of 45 K with a 90% settling time of 220 ms.Keywords
This publication has 5 references indexed in Scilit:
- Silicon nitride micromesh bolometer array for submillimeter astrophysicsApplied Optics, 2001
- SCUBA: a common-user submillimetre camera operating on the James Clerk Maxwell TelescopeMonthly Notices of the Royal Astronomical Society, 1999
- 100 mK bolometers for the submillimetre common-user bolometer array (scuba) I. Design and constructionInternational Journal of Infrared and Millimeter Waves, 1996
- Specific heat of nichrome, 2–30 KCryogenics, 1980
- Composite bolometers for submillimeter wavelengthsApplied Optics, 1978