Introduction: Lung protective ventilation requires low tidal volume and suitable positive end-expiratory pressure (PEEP). To date, few methods have been accepted for clinical use to set the appropriate PEEP. The aim of this study was to test the feasibility of PEEP titration guided by ventilation homogeneity using the global inhomogeneity (GI) index based on electrical impedance tomography (EIT) images. Methods: In a retrospective study, 10 anesthetized patients with healthy lungs mechanically ventilated under volume-controlled mode were investigated. Ventilation distribution was monitored by EIT. A standardized incremental PEEP trial (PEEP from 0 to 28 mbar, 2 mbar per step) was conducted. During the PEEP trial, "optimal" PEEP level for each patient was determined when the air was most homogeneously distributed in the lung, indicated by the lowest GI index value. Two published methods for setting PEEP were included for comparison based on the maximum global dynamic compliance and the intra-tidal compliance-volume curve. Results: No significant differences in the results were observed between the GI index method (12.2 ± 4.6 mbar) and the dynamic compliance method (11.4 ± 2.3 mbar, P > 0.6), or between the GI index and the compliance-volume curve method (12.2 ± 4.9 mbar, P > 0.6). Conclusions: According to the results, it is feasible and reasonable to use the GI index to select the PEEP level with respect to ventilation homogeneity. The GI index may provide new insights into the relationship between lung mechanics and tidal volume distribution and may be used to guide ventilator settings.Keywords
This publication has 33 references indexed in Scilit:
- A comparison of methods to identify open-lung PEEPIntensive Care Medicine, 2009
- Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomographyIntensive Care Medicine, 2007
- Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental studyCritical Care Medicine, 2007
- A Decremental PEEP Trial Identifies the PEEP Level That Maintains Oxygenation After Lung RecruitmentRespiratory Care, 2006
- Lung Volume Recruitment after Surfactant Administration Modifies Spatial Distribution of VentilationAmerican Journal of Respiratory and Critical Care Medicine, 2006
- Positive end‐expiratory pressure optimization using electric impedance tomography in morbidly obese patients during laparoscopic gastric bypass surgeryActa Anaesthesiologica Scandinavica, 2006
- Computed tomography scan assessment of lung volume and recruitment during high-frequency oscillatory ventilationCritical Care Medicine, 2005
- Regional Ventilation by Electrical Impedance TomographyChest, 2003
- An Objective Analysis of the Pressure-Volume Curve in the Acute Respiratory Distress SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Local mechanics of the lung tissue determined by functional EITPhysiological Measurement, 1996