Comparison of Three Junctional Tourniquets Using a Randomized Trial Design
Authors:
- Micah J. GasparyMGMicah J. Gaspary
- Gregory J. ZarowGZGregory J. Zarow
- Michael J. BarryMBMichael J. Barry
- Alexandra C. WalchakAWAlexandra C. Walchak
- Sean P. ConleySCSean P. Conley
- Paul J.D. RoszkoPRPaul J.D. Roszko
- 17 August 2018
- journal article
- research article
- Published by Taylor & Francis in Prehospital Emergency Care
- Vol. 23 (2) , 187-194
Abstract
Background: Hemorrhage remains a leading cause of death in both civilian and military settings. Of preventable deaths from hemorrhage, a significant portion occurs from junctional wounds that are not amenable to traditional extremity tourniquets. Junctional tourniquets (JTQs) can potentially provide hemorrhage control by compressing the arteries at the junction of the trunk and extremities. The FDA has cleared 3 JTQ products: The Combat Ready Clamp (CRoC®), the Junctional Emergency Treatment Tool (JETT™), and the SAM® Junctional Tourniquet (SJT). However, little is known regarding which of these JTQs is superior in application time, effectiveness in pulse elimination, effectiveness during transport, and user preference. Methods: Active duty corpsmen (N = 49) were given standardized instruction and hands-on training with the CRoC®, JETT™, and SJT, then sequentially applied each JTQ unilaterally to a fellow study participant in a randomized order. Pulse elimination was determined by Doppler ultrasound at the dorsalis pedis immediately then reevaluated after a short transport. User preference data were collected following testing. Data were analyzed using repeated measures ANOVA and non-parametric statistics at p < 0.05. Results: The CRoC® was significantly slower in application time than the JETT™ and SJT. Effectiveness was similar for CRoC®, JETT™, and SJT. Effectiveness during transport was significantly higher for SJT than for the JETT™, but no JTQ performed well during transport (24–48% effectiveness). SJT ranked first in perceived ease of use, stability, and reliability, and in user trust and overall preference. Participants provided cogent suggestions for product improvement. Conclusions: All JTQ devices performed poorly during transport. Combined, present findings highlight the potential of JTQ products for saving lives threatened by junctional wounds, but also highlight the need for specific product improvements towards fostering JTQ performance in patient transport.Keywords
Funding Information
- Deputy Surgeon General Clinical Investigation Award
This publication has 19 references indexed in Scilit:
- Testing of Junctional Tourniquets by Medics of the Israeli Defense Force in Control of Simulated Groin HemorrhageJournal of Special Operations Medicine, 2016
- Evaluation and Testing of Junctional Tourniquets by Special Operation Forces Personnel: A Comparison of the Combat Ready Clamp and the Junctional Emergency Treatment ToolJournal of Special Operations Medicine, 2016
- Quantification of residual limb skeletal muscle perfusion with contrast-enhanced ultrasound during application of a focal junctional tourniquetJournal of Vascular Surgery, 2016
- Performance of Junctional Tourniquets in Normal Human VolunteersPrehospital Emergency Care, 2014
- An analysis of prehospital deathsJournal of Trauma and Acute Care Surgery, 2014
- An Evidence-based Prehospital Guideline for External Hemorrhage Control: American College of Surgeons Committee on TraumaPrehospital Emergency Care, 2014
- In vivo assessment of the Combat Ready Clamp to control junctional hemorrhage in swineJournal of Trauma and Acute Care Surgery, 2013
- Death on the battlefield (2001–2011)Journal of Trauma and Acute Care Surgery, 2012
- Died of Wounds on the Battlefield: Causation and Implications for Improving Combat Casualty CareJournal of Trauma: Injury, Infection & Critical Care, 2011
- Magnetic resonance imaging measurement of blood volume flow in peripheral arteries in healthy subjectsJournal of Vascular Surgery, 2003