Piezoresistive torque magnetometry below 1 K

Authors:

Abstract
We have investigated the performance of piezoresistive cantilevers as magnetometers in the temperature range below 1 K. The de Haas-van Alphen effect was used to study the temperature dependence of a sample of κ-(BEDT-TTF)2Cu(NCS)2, fixed on the end of a cantilever, as a function of the excitation current through the piezoresistive device. We found that by using a small thermalizing wire connected directly to the sample, large excitations were not incompatible with sample temperatures remaining low, thereby establishing the use of these devices as sensitive magnetometers well below 1 K. A large hysteretic behavior observed at low fields (below 0.01 T) and low temperature (below ∼2 K) precludes their use in that regime. © 1999 American Institute of Physics

This publication has 7 references indexed in Scilit: