Optimal Approach for Small Bowel Capsule Endoscopy Using Polyethylene Glycol and Metoclopramide with the Assistance of a Real-Time Viewer
Authors:
- Hiroki EndoHEHiroki Endo
- Eiji SakaiESEiji Sakai
- Yusuke SekinoYSYusuke Sekino
- Takashi UchiyamaTUTakashi Uchiyama
- Seitaro WatanabeSWSeitaro Watanabe
- Hiroshi IidaHIHiroshi Iida
- Yasunari SakamotoYSYasunari Sakamoto
- Tomoko KoideTKTomoko Koide
- Hirokazu TakahashiHTHirokazu Takahashi
- Masato YonedaMYMasato Yoneda
- Chikako TokoroCTChikako Tokoro
- 15 April 2011
- journal article
- research article
- Published by S. Karger AG in Digestion
- Vol. 84 (2) , 119-125
Abstract
Aim: Capsule endoscopy is limited by the poor image quality of the distal bowel and incomplete small bowel transit. The aim of this study was to establish an optimal medication protocol for capsule endoscopy performed using a real-time viewer. Methods: A total of 80 patients were prospectively recruited. The patients were randomized into two groups: the ‘conventional group’ (without any preparation) and the ‘real-time group’ (in which a real-time viewer was attached). At 60 min after swallowing the capsule, if the capsule hadreached the small bowel, 500 ml of polyethylene glycol was administered; if the capsule was still located in the stomach, 10 mg of metoclopramide was given intramuscularly, followed by 500 ml of polyethylene glycol solution. Results: The completion rate was significantly higher in the real-time group as compared with that in the conventional group (72.5 vs. 90.0%). Our protocol yielded a significantly improved image quality of the distal small bowel [image quality score = 1.6 vs. 3.0 (max 4.0)]. The detection rate of lesions in the distal small bowel was higher in the real-time group than in the conventional group. Conclusions: The present study clearly showed that our protocol yielded an improved completion rate and also improved image quality.Keywords
This publication has 28 references indexed in Scilit:
- Optimal preparation for video capsule endoscopy: a prospective, randomized, single-blind studyEndoscopy, 2007
- A Single-Center Experience of 260 Consecutive Patients Undergoing Capsule Endoscopy for Obscure Gastrointestinal BleedingAmerican Journal of Gastroenterology, 2007
- Clinical Utility of Wireless Capsule EndoscopyJournal of Clinical Gastroenterology, 2006
- Complications, limitations, and failures of capsule endoscopy: a review of 733 casesGastrointestinal Endoscopy, 2005
- Is a 2-Liter PEG Preparation Useful Before Capsule Endoscopy?Journal of Clinical Gastroenterology, 2005
- Capsule endoscopy: role of bowel preparation in successful visualizationScandinavian Journal of Gastroenterology, 2004
- Capsule Endoscopy versus Push Enteroscopy in Patients with Occult Gastrointestinal BleedingZeitschrift für Gastroenterologie, 2003
- Wireless capsule video endoscopy is a superior diagnostic tool in comparison to barium follow-through and computerized tomography in patients with suspected Crohn's diseaseEuropean Journal of Gastroenterology & Hepatology, 2003
- Capsule endoscopy in the evaluation of patients with suspected small intestinal bleeding: Results of a pilot studyGastrointestinal Endoscopy, 2002
- Wireless-Capsule Diagnostic Endoscopy for Recurrent Small-Bowel BleedingNew England Journal of Medicine, 2001