A novel method for pulse transit time estimation using wrist bio-impedance sensing based on a regression model
Authors:
- 1 October 2017
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
Pulse transit time (PTT) is a promising method to estimate blood pressure using a wearable device. In this work, we explore the measurement of PTT using two bio-impedance (Bio-Z) sensors placed on the wrist 7.5 cm apart such that they can be integrated into a wrist watch. In order to take into account the different factors affecting blood flow in the wrist, a regression model is used based on multiple features extracted from the first and second derivative of Bio-Z. The estimated PTT is compared with a reference PTT measured from ECG and PPG sensors. Experimental measurements are conducted on three subjects and exhibited an average correlation coefficient of 0.77 for the PTT estimated from wrist Bio-Z compared to the reference PTT.Keywords
This publication has 5 references indexed in Scilit:
- Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A ReviewJournal of Medical Engineering, 2014
- A pulse transit time measurement method based on electrocardiography and bioimpedance2009
- A bio-impedance measurement system for portable monitoring of heart rate and pulse wave velocity using small body area2009
- Cuffless Estimation of Systolic Blood Pressure for Short Effort Bicycle Tests: The Prominent Role of the Pre-Ejection Period2006
- The valsalva maneuver as a test of cardiac functionThe American Journal of Medicine, 1957