Intraluminal micromanometry: an evaluation of the dynamic performance of micro‐extrusions and sleeve sensors
- 1 September 1996
- journal article
- Published by Wiley in Neurogastroenterology & Motility
- Vol. 8 (3) , 241-245
Abstract
Conventional manometric techniques are unsuitable for studies in premature infants and small laboratory animals. We have therefore developed silicone rubber 5‐lumen and 10‐lumen micromanometric extrusions with an o.d. 2.0 mm and lumina of 0.35 mm i.d. This study evaluates the suitability of micro‐extrusions for intraluminal perfusion manometry.Pressure offset, post‐occlusion pressure rise rate and sphincter model studies were used to assess the monometric performance of the extrusions and a miniature sleeve sensor (25 mm long) at infusion rates of 0.01‐0.1 mL min−1.Micro‐extrusions (5‐lumen/10‐lumen, respectively) had offsets (per 100 cm of length) of 3.8/5.0 mmHg at 0.01 mL min−1 and 25.6/26.2 mmHg at 0.1 mL min−1 and rise rates (in 160 cm lengths) of 64/43 mmHg sec−1 at 0.01 mL min−1 and 330/224 mmHg sec−1 at 0.1 mL min−1. Infusion rates 0.025 mL min−1 produced rise rates 100 mmHg sec−1. The miniature sleeve sensor had minimal resistance to perfusion, rise rates of 3 mmHg sec−1 at 0.01 mL min−1 and 23 mmHg sec−1 at 0.1 mL min−1 and recorded pressure as accurately as a side hole.We conclude that the performance of micromanometric extrusions and sleeves is sufficient for intraluminal perfusion manometry.Keywords
This publication has 6 references indexed in Scilit:
- Determinants of intrabolus pressure during esophageal peristaltic bolus transportAmerican Journal of Physiology-Gastrointestinal and Liver Physiology, 1993
- Sleeve device functions as a Starling resistor to record sphincter pressureAmerican Journal of Physiology-Gastrointestinal and Liver Physiology, 1985
- Improved Infusion System for Intraluminal Esophageal ManometryGastroenterology, 1977
- A New Technique for Continuous Sphincter Pressure MeasurementGastroenterology, 1976
- Factors Determining Pressure Measurement Accuracy by Intraluminal Esophageal ManometryGastroenterology, 1976
- Intraluminal Esophageal Manometry: An Analysis of Variables Affecting Recording Fidelity of Peristaltic PressuresGastroenterology, 1974