Home Discharge and Out-of-Hospital Follow-Up of Total Artificial Heart Patients Supported by a Portable Driver System

Authors:
  • LKLukacz Kizner
    Lukacz Kizner
  • LALatif Arusoglu
    Latif Arusoglu
  • MMMichael Morshuis
    Michael Morshuis
  • CBChristof Brehm
    Christof Brehm
  • RKReiner Koerfer
    Reiner Koerfer
  • CSChristoph Schuermann
    Christoph Schuermann
  • RSRichard G. Smith
    Richard G. Smith
  • JCJack G. Copeland
    Jack G. Copeland
  • MSMarvin J. Slepian
    Marvin J. Slepian

Abstract
To enhance ambulation and facilitate hospital discharge of total artificial heart (TAH)-supported patients, we adapted a mobile ventricular assistance device (VAD) driver (Excor) for TAH use and report on the performance of Excor-driven TAH patients discharged home. Ten patients stabilized on a TAH, driven by the CSS ("Circulatory Support System"), were progressively switched over to the Excor in hospital over 14 days as a pilot, with daily hemodynamics and laboratory parameters measured. Twenty-two stable TAH patients were subsequently placed on the Excor, trained, and discharged home. Clinical and hemodynamic parameters were followed. All pilot study patients were clinically stable on the Excor, with no decrease in TAH output noted (6.3 + 0.3 L/min [day 1] vs. 5.8 + 0.2 L/min [day 14], p = 0.174), with a trend suggesting improvement of both hepatic and renal function. Twenty-two TAH patients were subsequently successfully discharged home on the portable driver and were supported out of hospital for up to 598 days (range, 2-598; mean = 179 ± 140 days), remaining ambulatory, New York Heart Association (NYHA) class I or II, and free of readmission for 88.5% of the time of support. TAH patients may be effectively and safely supported by a mobile drive system. As such, the utility of the TAH may be extended to support patients beyond the hospital, at home, with overall ambulatory freedom.