Abstract
Objective: The SiPAP flow driver (Care Fusion, Dublin, Ohio, USA) offers synchronised nasal intermittent positive pressure ventilation (sNIPPV) using an abdominal capsule. This study aims to describe the accuracy and effects of synchronised NIPPV using SiPAP in preterm infants. Design: Ten infants, born 10% of breaths, resulting in asynchronous NIPPV delivery. Movements resulted in irregular SiPAP pressures and desaturation. When the RR was faster, >55/min, breaths irregularly triggered a SiPAP pressure peak (p=0.003). Compared with times when every breath resulted in a pressure peak, lower mean pressures were achieved, 7.9 vs 8.4 cm H2O (p=0.02). Conclusions: The SiPAP synchronisation system triggered rapidly with most spontaneous breaths, but did not result in larger tidal volumes. When the RR was >55/min, the SiPAP delivered fewer pressure peaks at lower pressures.