Injectable ultrasonic sensor for wireless monitoring of intracranial signals
Top Cited Papers
- 5 June 2024
- journal article
- research article
- Published by Springer Nature in Nature
- Vol. 630 (8015) , 84-90
Abstract
Direct and precise monitoring of intracranial physiology holds immense importance in delineating injuries, prognostication and averting disease1. Wired clinical instruments that use percutaneous leads are accurate but are susceptible to infection, patient mobility constraints and potential surgical complications during removal2. Wireless implantable devices provide greater operational freedom but include issues such as limited detection range, poor degradation and difficulty in size reduction in the human body3. Here we present an injectable, bioresorbable and wireless metastructured hydrogel (metagel) sensor for ultrasonic monitoring of intracranial signals. The metagel sensors are cubes 2 × 2 × 2 mm3 in size that encompass both biodegradable and stimulus-responsive hydrogels and periodically aligned air columns with a specific acoustic reflection spectrum. Implanted into intracranial space with a puncture needle, the metagel deforms in response to physiological environmental changes, causing peak frequency shifts of reflected ultrasound waves that can be wirelessly measured by an external ultrasound probe. The metagel sensor can independently detect intracranial pressure, temperature, pH and flow rate, realize a detection depth of 10 cm and almost fully degrade within 18 weeks. Animal experiments on rats and pigs indicate promising multiparametric sensing performances on a par with conventional non-resorbable wired clinical benchmarks.Keywords
This publication has 34 references indexed in Scilit:
- A wireless millimetre-scale implantable neural stimulator with ultrasonically powered bidirectional communicationNature Biomedical Engineering, 2020
- Biodegradable and flexible arterial-pulse sensor for the wireless monitoring of blood flowNature Biomedical Engineering, 2019
- Observation of elastic topological states in soft materialsNature Communications, 2018
- Invasive and noninvasive means of measuring intracranial pressure: a reviewPhysiological Measurement, 2017
- Small versus Large Catheters for Ventriculostomy in the Management of Intraventricular HemorrhageWorld Neurosurgery, 2017
- Bioresorbable silicon electronic sensors for the brainNature, 2016
- Continuous wireless pressure monitoring and mapping with ultra-small passive sensors for health monitoring and critical careNature Communications, 2014
- Sound and heat revolutions in phononicsNature, 2013
- Novel Methods to Predict Increased Intracranial Pressure During Intensive Care and Long-Term Neurologic Outcome After Traumatic Brain InjuryCritical Care Medicine, 2013
- CSF pressure assessed by lumbar puncture agrees with intracranial pressureNeurology, 2007