Modeling, characterization and test of a pediatric ventricular assist device
Open Access
- 18 June 2018
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 1044 (1) , 012047
Abstract
This paper presents a 0D model for the pulsatile pediatric Ventricular Assist Device (pVAD), as well as its characterization and test results obtained by using an experimental hydraulic system that provides operational conditions close to the physiological conditions observed in pediatric patients. The measurements of the physical quantities used in characterization of the pVAD have been obtained by using the LabVIEW platform. The air chamber pressure signal is used to model the pneumatic actuation system behaviour, whereas the pressure-volume relationship inside the blood chamber is applied to model the pVAD compliance. The set of test and validation results demonstrated the correctness of the proposed model.Keywords
This publication has 4 references indexed in Scilit:
- On deriving lumped models for blood flow and pressure in the systemic arteriesMathematical Biosciences and Engineering, 2004
- Characterization of an Adult Mock Circulation for Testing Cardiac Support DevicesASAIO Journal, 2004
- Avaliação do desempenho hemodinâmico do dispositivo de assistência ventricular InCor como substituto do coração esquerdoBrazilian Journal of Cardiovascular Surgery, 1999
- A modular numerical model of the cardiovascular system for studying and training in the field of cardiovascular physiopathologyJournal of Biomedical Engineering, 1992