Dosimetric comparison between cone/Iris‐based and InCise MLC‐based CyberKnife plans for single and multiple brain metastases
Authors:
Open Access
- 8 September 2016
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (5) , 184-199
Abstract
We performed an evaluation of the CyberKnife InCise MLC by comparing plan qualities for single and multiple brain lesions generated using the first version of InCise MLC, fixed cone, and Iris collimators. We also investigated differences in delivery efficiency among the three collimators. Twenty-four patients with single or multiple brain mets treated previously in our clinic on a CyberKnife M6 using cone/Iris collimators were selected for this study. Treatment plans were generated for all lesions using the InCise MLC. Number of monitor units, delivery time, target coverage, conformity index, and dose falloff were compared between MLC- and clinical cone/Iris-based plans. Statistical analysis was performed using the nonparametric Wilcoxon-Mann-Whitney signed-rank test. The planning accuracy of the MLC-based plans was validated using chamber and film measurements. The InCise MLC-based plans achieved mean dose and target coverage comparable to the cone/Iris-based plans. Although the conformity indices of the MLC-based plans were slightly higher than those of the cone/Iris-based plans, beam delivery time for the MLC-based plans was shorter by . For smaller targets or cases with OARs located close to or abutting target volumes, MLC-based plans provided inferior dose conformity compared to cone/Iris-based plans. The QA results of MLC-based plans were within 5% absolute dose difference with over 90% gamma passing rate using gamma criteria. The first version of InCise MLC could be a useful delivery modality, especially for clinical situations for which delivery time is a limiting factor or for multitarget cases. PACS number(s): 87.53.Ly, 87.55.D-
Keywords
This publication has 28 references indexed in Scilit:
- Variable Circular Collimator in Robotic Radiosurgery: A Time-Efficient Alternative to a Mini-Multileaf Collimator?International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Stereotactic, Single-Dose Irradiation of Lung Tumors: A Comparison of Absolute Dose and Dose Distribution Between Pencil Beam and Monte Carlo Algorithms Based on Actual Patient CT ScansInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Stereotactic Radiosurgery for the Management of Brain MetastasesNew England Journal of Medicine, 2010
- Radiotherapy Treatment Plans With RapidArc for Prostate Cancer Involving Seminal Vesicles and Lymph NodesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- The role of stereotactic radiosurgery in the management of patients with newly diagnosed brain metastases: a systematic review and evidence-based clinical practice guidelineJournal of Neuro-Oncology, 2009
- Volumetric Modulated Arc Therapy for Delivery of Prostate Radiotherapy: Comparison With Intensity-Modulated Radiotherapy and Three-Dimensional Conformal RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Cyberknife Image-Guided Delivery and Quality AssuranceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- Radiation therapy oncology group: Radiosurgery quality assurance guidelinesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1993