Biomechanical Reactions of Exoskeleton Neurorehabilitation Robots in Spastic Elbows and Wrists
Authors:
- 9 June 2017
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 25 (11) , 2196-2203
Abstract
Spasticity is an important factor in designing wearable and lightweight exoskeleton neurorehabilitation robots. The proposed study evaluates biomechanical reactions of an exoskeleton robot to spasticity and establishes relevant guidelines for robot design. A two-axis exoskeleton robot is used to evaluate a group of 20 patients post-stroke with spastic elbow and/or wrist joints. All subjects are given isokinetic movements at various angular velocities within the capable range of motion for both joints. The resistance torque and corresponding angular position at each joint are recorded continuously. Maximal resistance torques caused by low (modified Ashworth scale (MAS) 0, 1), intermediate (MAS 1+), and high (MAS 2 and 3) grade spasticity were 3.68 ± 2.42, 5.94 ± 2.55, and 8.25 ± 3.35 Nm for the elbow flexor (p <; 0.001, between each grades) and 4.23 ± 1.75, 5.68 ± 1.96, and 5.44 ± 2.02 Nm for the wrist flexor (p <; 0.001, for low versus intermediate, low versus high grade spasticity). The angular velocity did not significantly influence maximal resistance torque in either joint. The catch occurred more quickly at higher velocities for low and intermediate elbow flexor spasticity (p <; 0.01). Spasticity caused considerable resistance to the robots during mechanically actuated movements. The resistance range according to the degree of spasticity should be considered when designing practical neurorehabilitation robots.Keywords
Funding Information
- Brain Fusion Program of Seoul National University (800-20120444)
This publication has 29 references indexed in Scilit:
- Neurophysiology of Robot-Mediated Training and Therapy: A Perspective for Future Use in Clinical PopulationsFrontiers in Neurology, 2013
- Spasticity After StrokeAmerican Journal of Physical Medicine & Rehabilitation, 2012
- Neuro-rehabilitation robot-assisted assessments of synergy patterns of forearm, elbow and shoulder joints in chronic stroke patientsClinical Biomechanics, 2010
- Characterization of spasticity in cerebral palsy: dependence of catch angle on velocityDevelopmental Medicine & Child Neurology, 2010
- An investigation into the agreement between clinical, biomechanical and neurophysiological measures of spasticityClinical Rehabilitation, 2008
- The relation between Ashworth scores and neuromechanical measurements of spasticity following strokeJournal of NeuroEngineering and Rehabilitation, 2008
- Biomechanical measurement of post-stroke spasticityAge and Ageing, 2006
- Reliability of Biomechanical Spasticity Measurements at the Elbow of People PoststrokeArchives of Physical Medicine and Rehabilitation, 2005
- Spasticity after spinal cord injurySpinal Cord, 2005
- Adjustments to Zatsiorsky-Seluyanov's segment inertia parametersJournal of Biomechanics, 1996