A low-temperature scanning tunneling microscope capable of microscopy and spectroscopy in a Bitter magnet at up to 34 T
Open Access
- 1 September 2017
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 88 (9) , 093706
Abstract
We present the design and performance of a cryogenic scanning tunneling microscope (STM) which operates inside a water-cooled Bitter magnet, which can attain a magnetic field of up to 38 T. Due to the high vibration environment generated by the magnet cooling water, a uniquely designed STM and a vibration damping system are required. The STM scan head is designed to be as compact and rigid as possible, to minimize the effect of vibrational noise as well as fit the size constraints of the Bitter magnet. The STM uses a differential screw mechanism for coarse tip—sample approach, and operates in helium exchange gas at cryogenic temperatures. The reliability and performance of the STM are demonstrated through topographic imaging and scanning tunneling spectroscopy on highly oriented pyrolytic graphite at T = 4.2 K and in magnetic fields up to 34 T.Keywords
All Related Versions
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (VIDI 680-47-534)
- Stichting voor Fundamenteel Onderzoek der Materie
This publication has 25 references indexed in Scilit:
- Hofstadter’s butterfly and the fractal quantum Hall effect in moiré superlatticesNature, 2013
- 18/20 T high magnetic field scanning tunneling microscope with fully low voltage operability, high current resolution, and large scale searching abilityReview of Scientific Instruments, 2012
- Scanning Tunneling Spectroscopy of Graphene on GraphitePhysical Review Letters, 2009
- Landau levels and oscillator strength in a biased bilayer of graphenePhysical Review B, 2007
- Observation of Landau levels of Dirac fermions in graphiteNature Physics, 2007
- Room-Temperature Quantum Hall Effect in GrapheneScience, 2007
- Real-Space Imaging of Alternate Localization and Extension of Quasi-Two-Dimensional Electronic States at Graphite Surfaces in Magnetic FieldsPhysical Review Letters, 2006
- A low temperature scanning tunneling microscope for use in high magnetic fieldsReview of Scientific Instruments, 1995
- Protein secondary structures in water from second-derivative amide I infrared spectraBiochemistry, 1990
- 7 × 7 Reconstruction on Si(111) Resolved in Real SpacePhysical Review Letters, 1983