Effect of Alteration in Pump Speed on Pump Output and Left Ventricular Filling with Continuous-Flow Left Ventricular Assist Device
- 1 November 2011
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 57 (6) , 495-500
Abstract
Third-generation continuous-flow left ventricular assist devices (LVAD) provide reduced pulsatility flow. We examined the safe working range for LVAD pump speed and the effect on pump output and cardiac function in 13 stable outpatients with VentrAssist-LVAD (Ventracor Ltd, Australia). Pump speed was decreased from a baseline mean of 2,073 ± 86 revolutions per minute (RPM, with corresponding mean flow of 5.59 ± 1.18 L/min, mean ± standard deviation) to an average low-speed of 1,835 ± 55 RPM (corresponding flow 4.68 ± 0.99 L/min) and up to high-speed of 2,315 ± 66 RPM (corresponding flow 6.30 ± 1.29 L/min). There was a strong linear relationship between alteration in speed and flow rates (r2 = 0.89, p < 0.00001) but marked interpatient variation. Downward titration to preset minimum 1,800 RPM was achieved in 9/13 (69%) and upward titration to the preset maximum 2,400 RPM was achieved in 4/13 (31%). Upward titration was stopped due to ventricular suction or nonsustained ventricular tachycardia (VT) in 4/13 (31%). Ventricular suction or VT (in 4/13) tended to be more common in patients with poor right ventricular (RV) function (p = 0.07). In summary, pump flow is stable within a relatively small speed range and should not be altered without close monitoring due to variation in response between patients, particularly with concomitant RV impairment.Keywords
This publication has 15 references indexed in Scilit:
- Effects of the HeartMate II continuous-flow left ventricular assist device on right ventricular functionThe Journal of Heart and Lung Transplantation, 2010
- Second INTERMACS annual report: More than 1,000 primary left ventricular assist device implantsThe Journal of Heart and Lung Transplantation, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist DeviceJACC, 2009
- A Prospective, Multicenter Trial of the VentrAssist Left Ventricular Assist Device for Bridge to Transplant: Safety and EfficacyThe Journal of Heart and Lung Transplantation, 2008
- Initial Clinical Experience With the VentrAssist Left Ventricular Assist Device: The Pilot TrialThe Journal of Heart and Lung Transplantation, 2008
- Suction Due to Left Ventricular Assist: Implications for Device Control and ManagementArtificial Organs, 2007
- Hemodynamic Controller for Left Ventricular Assist Device Based on Pulsatility RatioArtificial Organs, 2007
- Statin Attenuates Increase in C-Reactive Protein During Estrogen Replacement Therapy in Postmenopausal WomenCirculation, 2002
- Flow characteristics and required control algorithm of an implantable centrifugal left ventricular assist deviceHeart and Vessels, 1997