Determination of Serum Triglycerides by Capillary On-Column Gas Chromatography
Authors:
- J. W. I. BrunnekreeftJBJ. W. I. Brunnekreeft
- B. LeijnseBLB. Leijnse
- 1 January 1986
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 24 (7) , 445-450
Abstract
We report an accurate method, based on capillary on-column gas chromatography, for determining triglycerides in human serum. After serum extraction and chemical hydrolysis, glycerol is directly measured by gas chromatography (GC). With our extraction method no free glycerol is extracted from serum. The accuracy of this method was compared with that of a method based on the original procedure of Carlson (J. Atheroscler. Res. 3, 334-336 (1963)) and which is standardized by the Centers for Disease Control (CDC, Atlanta, U.S.A.). Orthogonal regression analysis of the GC method (y) and the CDC reference method (x) resulted in y = 0.996 x + 0.000 with a correlation coefficient of 0.999. The variances of analytical data, collected over a two month period showed that, for the GC method, the within-day coefficient of variation (CV) was < 1.43%; the between-day CV < 1.35%. The data for the DCD method were < 3.36% (within-day) and < 6.38% (between-days). The CDC method is linear up to 3.7 mmol/l and the GC method to 22 mmol/l.Keywords
This publication has 1 reference indexed in Scilit:
- Determination of serum triglyceridesJournal of Atherosclerosis Research, 1963