Abstract
Screen printed electrodes were employed in a proof of concept determination of homocysteine and glutathione using electrochemically oxidized catechol via a 1,4-Michael addition reaction in the absence and presence of cysteine, and each other. Using cyclic voltammetry, the Michael reaction introduces a new adduct peak which is analytically useful in detecting thiols. The proposed procedure relies on the different rates of reaction of glutathione and homocysteine with oxidized catechol so that at fast voltage scan rates only homocysteine is detected in cyclic voltammetry. At slower scan rates, both glutathione and homocysteine are detected. The combination of the two sets of data provides quantification for homocysteine and glutathione. The presence of cysteine is shown not to interfere provided sufficient high concentrations of catechol are used. Calibration curves were determined for each homocysteine and glutathione detection; where the sensitivities are 0.019 µA·µM−1 and 0.0019 µA·µM−1 and limit of detections are ca. 1.2 µM and 0.11 µM for homocysteine and glutathione, respectively, within the linear range. This work presents results with potential and beneficial use in re-useable and/or disposable point-of-use sensors for biological and medical applications.Keywords
This publication has 55 references indexed in Scilit:
- Electrochemical determination of glutathione: a reviewThe Analyst, 2012
- Hg2+ detection by measuring thiol groups with a highly sensitive screen-printed electrode modified with a nanostructured carbon black filmElectrochimica Acta, 2011
- Simultaneous electrochemical determination of glutathione and tryptophan on a nano-TiO2/ferrocene carboxylic acid modified carbon paste electrodeSensors and Actuators B: Chemical, 2009
- Electrochemical Reduction of Ellman's Reagent: A Novel Selective Detection Protocol for Thiol CompoundsElectroanalysis, 2007
- Electroanalytical Exploitation of Nitroso Phenyl Modified Carbon‐Thiol Interactions: Application to the Low Voltage Determination of ThiolsElectroanalysis, 2007
- Electrochemical Detection of Glutathione Using Redox IndicatorsAnalytical Chemistry, 2006
- An electrochemical adaptation of Ellman’s testThe Analyst, 2002
- Electrochemically initiated 1,4 additions: a versatile route to the determination of thiolsAnalytica Chimica Acta, 2001
- Glutathione and cysteine measurement in biological samples by HPLC with a glassy carbon working detectorBiomedical Chromatography, 1994
- Determination of reduced glutathione in guinea pig and rat tissue by HPLC with electrochemical detectionLife Sciences, 1978