Bedside monitoring of CBF with xenon-CT and a mobile scanner: a novel method in neurointensive care
Authors:
- J. HillmanJHJ. Hillman
- P. SturnegkPSP. Sturnegk
- H. YonasHYH. Yonas
- J. HeronJHJ. Heron
- M. SandborgMSM. Sandborg
- T. GunnarssonTGT. Gunnarsson
- P. MellergårdPMP. Mellergård
- 1 January 2005
- journal article
- case report
- Published by Taylor & Francis in British Journal Of Neurosurgery
- Vol. 19 (5) , 395-401
Abstract
Combining previously independently established techniques our objective was to develop and evaluate a method for bedside qualitative assessment of cerebral blood flow in neurointensive care (NICU) patients. The CT-protocol was optimized using phantoms and comparing a mobile CT-scanner (Tomoscan-M, Philips) with two stationary CT scanners. Thirty-two per cent xenon was delivered with standard equipment (Enhancer 3000). Mean cortical flow in volunteers was 48 ml/min/100 g, with the mean vascular territorial flow varying between 45 and 66 ml/min/100 g. The potential clinical usefulness was illustrated in three patients with vasospasm following subarachnoid haemorrhage. Our conclusion is that quantitative bedside measurements of CBF can be repeatedly performed in an easy and safe way in a standard NICU-setting, using xenon-inhalation and a mobile CT-scanner. The method is useful for the decision-making, and is a good example of how the quality of multi-modality monitoring in the NICU can be developed and further diversified.Keywords
This publication has 19 references indexed in Scilit:
- Jugular Venous Oxygen Saturation Thresholds in Trauma Patients May Not Extrapolate to Ischemic Stroke PatientsJournal of Neurosurgical Anesthesiology, 2002
- CT-Based Assessment of Acute Stroke: CT, CT Angiography, and Xenon-Enhanced CT Cerebral Blood FlowStroke, 2001
- Mobile computerized tomography scanning in the neurosurgery intensive care unit: increase in patient safety and reduction of staff workloadJournal of Neurosurgery, 2000
- Effect of Xenon-Induced Flow Activation on Xenon-Enhanced Computed Tomography Cerebral Blood Flow CalculationsJournal of Cerebral Blood Flow & Metabolism, 1998
- CBF Measured by Xe-CT: Approach to Analysis and Normal ValuesJournal of Cerebral Blood Flow & Metabolism, 1991
- Stable Xenon Does Not Increase Intracranial Pressure in Primates with Freeze-Injury-Induced Intracranial HypertensionJournal of Cerebral Blood Flow & Metabolism, 1991
- Tomographic cerebral blood flow measurement using xenon inhalation and serial CT scanning: Normal values and its validityNeurosurgical Review, 1985
- Mapping local blood flow of human brain by CT scanning during stable xenon inhalation.Stroke, 1981
- Movement of criticall ill patients within hospital.BMJ, 1975
- Regional Cerebral Blood Flow Estimated by 133 Xenon InhalationStroke, 1975