Influence of the plain-parallel electrode surface dimensions on the type A measurement uncertainty of GM counter
Author:
Open Access
- 1 January 2011
- journal article
- Published by National Library of Serbia in Nuclear Technology and Radiation Protection
- Vol. 26 (1) , 39-44
Abstract
This paper investigates, through theory and experiment, the influence of the plain-parallel electrode surface dimensions change on the type A measurement uncertainty of a GM counter. The possibilities of applying these results to practical structures are examined by using the methods of mathematical statistics. Special attention is devoted to the influence of electrode surface enlargement on the statistical behavior of the pulse number random variable, expressed in the form of the enlargement law. In the theoretical part of the paper, the general surface enlargement law is derived. Comparison of experimental results with those predicted by the surface enlargement law proved its validity for expressing the type A measurement uncertainty of GM counters constructed with a plain-parallel electrode configuration with a homogenous electric field.Keywords
This publication has 6 references indexed in Scilit:
- Validity of the Space-Time Enlargement Law for vacuum breakdownVacuum, 2010
- Influence of tube volume on measurement uncertainty of GM countersNuclear Technology and Radiation Protection, 2010
- Time enlargement law for gas pulse breakdownPlasma Sources Science and Technology, 2009
- Reliability of semiconductor and gas-filled diodes for over-voltage protection exposed to ionizing radiationNuclear Technology and Radiation Protection, 2009
- Surface-time enlargement law for gas breakdownIEEE Transactions on Dielectrics and Electrical Insulation, 2008
- Influence of radiation energy and angle of incidence on the uncertainty in measurements by GM countersNuclear Technology and Radiation Protection, 2008