Embolization of an Amplatzer Atrial Septal Closure Device to the Pulmonary Artery
Authors:
- 1 March 2008
- journal article
- case report
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 23 (2) , 164-167
Abstract
A 44-year-old woman with a history of transient ischemic attack underwent closure of atrial septal defect with a 26 mm Amplatzer device. The device was released without residual shunt or impingement on intracardiac structures. Within seconds, the transesophageal echocardiography showed the initial dislodgement of the device from the atrial septum and its consequent slipping back into the right atrium close to the tricuspid valve. Soon after the device disappeared from the right atrium and it could be founded into the right ventricle under the tricuspid valve. The patient was transferred in the operating room for an emergency operation. The device could not be found in the right ventricle because its downstream migration. The Amplatzer septal occluder was identified by palpation into the pulmonary artery trunk: it was retrieved from the right ventricle through the pulmonary valve and the atrial septal defect was closed by running suture.Keywords
This publication has 14 references indexed in Scilit:
- Perforation of the right atrium and the ascending aorta following percutaneous transcatheter atrial septal defect closureInteractive CardioVascular and Thoracic Surgery, 2005
- Late cardiac perforation following transcatheter atrial septal defect closureThe Annals of Thoracic Surgery, 2004
- Surgical management of complications after transcatheter closure of an atrial septal defect or patent foramen ovaleThe Journal of Thoracic and Cardiovascular Surgery, 2000
- Percutaneous Closure of Patent Foramen Ovale in Patients With Paradoxical EmbolismCirculation, 2000
- Operative findings in failed button device closure of ASDThe Annals of Thoracic Surgery, 1995
- International experience with secundum atrial septal defect occlusion by the buttoned deviceAmerican Heart Journal, 1994
- Transcatheter closure of interatrial communications with a modified umbrella device.Heart, 1994
- Residual shunts after transcatheter closure of patent ductus arteriosus. A major concern or benign "techno-malady"?Circulation, 1991
- Double-umbrella closure of atrial defects. Initial clinical applications.Circulation, 1990
- Secundum atrial septal defect. Nonoperative closure during cardiac catheterizationJAMA, 1976