Performance of EasyBreath Decathlon Snorkeling mask for delivering continuous positive airway pressure
Open Access
- 10 March 2021
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 11 (1) , 1-8
Abstract
During the COVID-19 pandemic, the need for noninvasive respiratory support devices has dramatically increased, sometimes exceeding hospital capacity. The full-face Decathlon snorkeling mask, EasyBreath (EB mask), has been adapted to deliver continuous positive airway pressure (CPAP) as an emergency respiratory interface. We aimed to assess the performance of this modified EB mask and to test its use during different gas mixture supplies. CPAP set at 5, 10, and 15 cmH2O was delivered to 10 healthy volunteers with a high-flow system generator set at 40, 80, and 120 L min−1 and with a turbine-driven ventilator during both spontaneous and loaded (resistor) breathing. Inspiratory CO2 partial pressure (PiCO2), pressure inside the mask, breathing pattern and electrical activity of the diaphragm (EAdi) were measured at all combinations of CPAP/flows delivered, with and without the resistor. Using the high-flow generator set at 40 L min−1, the PiCO2 significantly increased and the system was unable to maintain the target CPAP of 10 and 15 cmH2O and a stable pressure within the respiratory cycle; conversely, the turbine-driven ventilator did. EAdi significantly increased with flow rates of 40 and 80 L min−1 but not at 120 L min−1 and with the turbine-driven ventilator. EB mask can be safely used to deliver CPAP only under strict constraints, using either a high-flow generator at a flow rate greater than 80 L min−1, or a high-performance turbine-driven ventilator.Keywords
All Related Versions
This publication has 25 references indexed in Scilit:
- Noninvasive CPAP With Face Mask: Comparison Among New Air-Entrainment Masks and the Boussignac ValveRespiratory Care, 2013
- How are medical devices regulated in the European Union?Journal of the Royal Society of Medicine, 2012
- Assessing the performance of the Whisperflow® continuous positive airway pressure generator: a bench studyBritish Journal of Anaesthesia, 2009
- Physiological effects of different interfaces during noninvasive ventilation for acute respiratory failure*Critical Care Medicine, 2009
- Introducing new anaesthetic equipment into clinical practiceAnaesthesia, 2008
- Helmet ventilation and carbon dioxide rebreathing: effects of adding a leak at the helmet portsIntensive Care Medicine, 2008
- Electrical Activity of the Diaphragm during Pressure Support Ventilation in Acute Respiratory FailureAmerican Journal of Respiratory and Critical Care Medicine, 2001
- Effects of Periodic Positive Airway Pressure by Mask on Postoperative Pulmonary FunctionChest, 1986
- Treatment of the Adult Respiratory Distress Syndrome with Continuous Positive Airway PressureChest, 1979
- Mechanical Work on the Lungs and Work of Breathing with Positive End-Expiratory Pressure and Continuous Positive Airway PressureChest, 1979