Importance of Ventricular Rate After Mode Switching During Low Intensity Exercise as Assessed by Clinical Symptoms and Ventilatory Gas Exchange
- 1 January 2000
- journal article
- clinical trial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 23 (1) , 32-39
Abstract
Automatic mode switching from DDD(R) to DDI(R) or VVI(R) pacing modes has improved dual chamber pacing in patients at high risk for supraventricular tachyarrhythmias. However, little is known about the effect of ventricular pacing rate adaptation after mode switching. We conducted a single-blinded, crossover study in 15 patients (58 +/- 21 years) with a DDD pacemaker who had AV block and normal sinus node function to investigate the influence of pacing rate adaptation to intrinsic heart rate during low intensity exercise. Patients performed two tests (A/B) of low intensity treadmill exercise (0.5 W/kg) in randomized order. They initially walked for 6 minutes while paced in DDD mode. The pacing mode was then switched to VVI with a pacing rate of either 70 beats/min (test A) or matched to the intrinsic heart rate (95 +/- 11 beats/min test B). Respiratory gas exchange variables were determined and patients classified the effort before and after mode switching on a Borg scale from 6 to 20. Percentage changes of respiratory gas exchange measurements were significantly larger (O2 consumption: -8.2 +/- 5.0% vs. -0.6 +/- 7.2%; ventilatory equivalent of CO2 exhalation: 5.3 +/- 4.9% vs. 1.5 +/- 4.3%; respiratory exchange ratio: 7.0 +/- 2.2% vs. 3.5 +/- 3.0%; end-tidal CO2: -5.7 +/- 2.9% vs. -1.8 +/- 2.7%; all P < 0.01) and the increase in subjective assessment of the effort tended to be higher (mean increase on Borg scale: 1.6 +/- 1.9 vs. 1.1 +/- 1.8, P = 0.07) after heart rate unadjusted than after adjusted mode switching. Mode switching from DDD to VVI pacing is better tolerated and gas exchange measurements are less influenced if ventricular pacing rate is adjusted to the level of physical activity. Thus, pacing rate adjustment should be considered as part of automatic mode switch algorithms.Keywords
This publication has 27 references indexed in Scilit:
- Risk factors for the development of recurrent atrial fibrillation: Role of pacing and clinical variablesAmerican Heart Journal, 1995
- Cardiac output in single-lead VDD pacing versus rate-matched VVIR pacingThe American Journal of Cardiology, 1995
- Hemodynamics of Idiopathic Paroxysmal Atrial FibrillationPacing and Clinical Electrophysiology, 1995
- Quality‐of‐Life in DDDR Pacing: Atrioventricular Synchrony or Rate Adaptation?Pacing and Clinical Electrophysiology, 1994
- Single- versus dual-chamber sensor-driven pacing: Comparison of cardiac outputsAmerican Heart Journal, 1991
- Pacemaker Syndrome Evaluated by Cardiopulmonary Exercise TestingPacing and Clinical Electrophysiology, 1990
- Benefits of physiologic atrioventricular synchronization for pacing with an exercise rate responseThe American Journal of Cardiology, 1990
- Single Versus Double Chamber Rate Responsive Cardiac Pacing: Comparison by Cardiopulmonary Noninvasive Exercise TestingPacing and Clinical Electrophysiology, 1988
- Clinical and hemodynamic comparison of VVI versus DDD pacing in patients with DDD pacemakersThe American Journal of Cardiology, 1988
- Hemodynamic Effects of Atrial Synchronization in Acute and Long‐Term Ventricular PacingPacing and Clinical Electrophysiology, 1982