Ambient-Temperature Trap/Release of Arsenic by Dielectric Barrier Discharge and Its Application to Ultratrace Arsenic Determination in Surface Water Followed by Atomic Fluorescence Spectrometry

Authors:

Abstract
A novel dielectric barrier discharge reactor (DBDR) was utilized to trap/release arsenic coupled to hydride generation atomic fluorescence spectrometry (HG-AFS). On the DBD principle, the precise and accurate control of trap/release procedures was fulfilled at ambient temperature, and an analytical method was established for ultratrace arsenic in real samples. Moreover, the effects of voltage, oxygen, hydrogen and water vapor on trapping and releasing arsenic by DBDR were investigated. For trapping, arsenic could be completely trapped in DBDR at 40 mL/min of O2 input mixed with 600 mL/min Ar carrier gas and 9.2 kV discharge potential; prior to release, the Ar carrier gas input should be changed from the upstream gas liquid separator (GLS) to the downstream GLS and kept for 180 s to eliminate possible water vapor interferences; for arsenic release, O2 was replaced by 200 mL/min H2 and discharge potential was adjusted to 9.5 kV. Under optimized conditions, arsenic could be detected as low as 1.0 ng/L with a 8-fold enrichment factor; the linearity of calibration ranged from 0.05 μg/L to 5 μg/L with a R2 > 0.995. The mean spiked recoveries for tap, river, lake and sea water samples were 98% to 103%; and the measured values of the CRMs including GSB-Z50004-200431, GBW08605, and GBW(E)080390 were in good agreement with the certified values. These findings prove the feasibility of DBDR as an arsenic pre-concentration tool for atomic spectrometric instrumentation and arsenic recycle in industrial waste gas discharge.
Funding Information
  • Ministry of Science and Technology of the People's Republic of China (2011YQ140149)
  • National Natural Science Foundation of China (31301491, 31571924)

This publication has 30 references indexed in Scilit: